![Emerson 6081-P Скачать руководство пользователя страница 9](http://html.mh-extra.com/html/emerson/6081-p/6081-p_instruction-manual_104604009.webp)
Description and Specifications
1
Instruction Manual
Description and Specifications
LIQ_MAN_6081-P
February 2014
Section 1: Description and Specifications
• High accuracy and reliability for monitoring applications
• Self-organizing network for high data reliability and network stability
• Industry Leading wireless Security
• Compatible with 1420 Wireless gateway and Emerson Process Management WirelessHART
TM
networks
• Easy to read two-line display with easy to use menus
• WirelessHART 7 Digital Communications
• SMART Sensor Enabled
• Continuous diagnostics monitor sensor performance and health
1.1 Description and Specifications
The 6081-P transmitter is ideal for monitoring applications, especially in hard-to-reach or cost-
prohibitive locations. The 6081-P measures pH and ORP and is compatible with most Rosemount
Analytical pH and ORP sensors. The transmitter has a rugged, cast aluminum weatherproof
and corrosion-resistant enclosure (NEMA 4X). The transmitter includes a two-line 16-character
display with simple and intuitive menu screens. Plain language prompts in six (6) local languages
guide the user through the programming and calibration procedures. The 6081 is compatible
with non-preamp pH and ORP sensors and SMART pH sensors from Rosemount Analytical.
Installation and start-up of the 6081-P wireless transmitter is simple. Just power the
6081-P and assign it to a wireless network with a 1420 Gateway. The unit will auto-locate the
most efficient path to the host and will begin transmitting measurement data immediately via 2.4
GHz wireless communications. The Self-Organizing Network ensures exceptional data reliability
and network stability. All of Emerson Process Management’s wireless devices employ Encryption,
Authentication, Verification, Anti-Jamming and Key Management to ensure data transmission and
security. Rosemount Analytical devices include intelligent power management to reduce power
consumption and extend power module life while delivering highly reliable measurements with
rich HART data and diagnostic information. HART digital communication allows access to AMS
(Asset Management Solutions) for live process variables, useful diagnostics and troubleshooting
information.
1.2 Specifications - General
Enclosure: Cast aluminum. NEMA 4X.
Dimensions: 6.55” x 5.40” x 5.15” (166mm x 137mm x 131mm)
Conduit Openings: 3/4” FNPT
Ambient Temperature: 32 to 122°F (0 to 50°C)
Storage Temperature: -4 to 158°F (-20 to 70°C)
Relative Humidity: 0 to 95% (non-condensing)
Weight/Shipping Weight: 7 lbs/8 lbs (3.2/3.6 kg)
RFI/EMI: EN-61326
Digital Communications: HART 7 WirelessHART
1.2.1 Specifications - Wireless
Output: Wireless enabled HART 7.0
Transmit Rate: User selectable.
Antenna: PBT/PC integrated omni-directional antenna
1.3 Specifications - Wireless
Output:
WirelessHART V7
Transmit Rate:
User selectable, 1/sec. to 1/60 min (via 1420 wireless Gateway or AMS)
Measurement update rate:
1/sec. to
1/10
min
Antenna:
PBT/PC integrated omni-directional antenna