Emec J DIGITAL CL Скачать руководство пользователя страница 8

8

CALIBRATION:  2)CALIB

Choose “CALIB” from main menu. Display shows:

Calib

1)Zero

Press “UP” key. Display shows:

Calib

2) Slope

Zero calibration.

When display shows “1) Zero” press “E” to enter into zero calbration procedure. Display shows:

R 1.00Cl

C 0.00Cl

To  calibrate  it  ,  if  you  are  using  an  ECL4-6-7  probe,  close  water’s  flow  and  wait  for  some  minutes. 
If you are using an ECL1-2-3-8 probe use flow-water without chlorine into probe holder or install an 
activated-carbon cartridge. If no one of these methods is vailable, unscrew top probe connector or, if 
not present, disconnect probe’s green wire (signal wire). Wait until “R” value is stable. ATTENTION: IT 

IS NOT NECESSARY  THAT “R”  VALUE WILL BE THE SAME OF “C” VALUE. Press “E” to confirm. 

Display shows:

Calibr.

Success.

Slope calibration.

When display shows “2) Slope” press “E” to enter into slope calbration procedure. Display shows:

R 1.00Cl

C 0.00Cl

To obtain reliable reading results is necessary to calibrate the probe using a chlorine solution. Verify the 

plant’s chlorine value at  the end of probe holder using a colorimetric (DPD1) analysis sistem. Using 

“UP” and “DOWN” keys enter this value in “C” field. Wait until “R” value is stable. ATTENTION: IT IS 

NOT NECESSARY  THAT “R”  VALUE WILL BE THE SAME OF “C” VALUE. 

SETUP

Содержание J DIGITAL CL

Страница 1: ...ns important safety informations about installation and use of this equipment Ignoring this informations could result in injuries or damages It is strictly forbidden to use this equipment with radioac...

Страница 2: ...authorized personnel only Always discharge the liquid end before servicing the instrument Empty and rinse the liquid end before work on a pump which has been used with hazardous or unknown chemicals A...

Страница 3: ...VIEW ON MENU 6 SETTING SETPOINTS 1 SETPN 6 Out mA 7 CALIBRATION 2 CALIB 8 DELAY 3 Delay 9 RESERVED 4 Reserv 9 PASSWORD SETUP 1 New Pw 10 STANDBY SETUP 2 Stand 10 MANUAL WORKING MODE 3 Serv 11 ERROR ME...

Страница 4: ...red data The Enter key confirms the selection Press RIGHT key for about 4 seconds to switch off the instrument the display shows OFF Repeat the sequence to return to the normal operating mode To reset...

Страница 5: ...instrument is set to work without a proximity sensor If needed proximity sensor turns the instrument into stand by mode if there isn t flow Instrument s display shows NO FLOW Connect SEPR to blocks 13...

Страница 6: ...MAIN menu is configured as follows 1 Setup This menu is divided into 1 Setpn 2 Calib 3 Delay 2 Param This menu is divided into 1 New Pw 2 Stand 3 Serv service mode SETTING SETPOINTS 1 SETPN Out 1 and...

Страница 7: ...and OFF It is not possible to enter mixed values ex 100 for SP1 and ON for SP2 In this case it will not possible save them In proportional mode the instrument doesn t show values for setpoints status...

Страница 8: ...top probe connector or if not present disconnect probe s green wire signal wire Wait until R value is stable ATTENTION IT IS NOT NECESSARY THAT R VALUE WILL BE THE SAME OF C VALUE Press E to confirm...

Страница 9: ...aches the setpoint values Default value is set to 0 Select DELAY from SETUP menu The instrument will show OUT1 10 Sec Press RIGHT key The instrument will show OUT2 5 Sec Delay time can be set from 0 n...

Страница 10: ...Press E to confirm The instrument will show the new password for 2 seconds and then will return to the main menu STANDBY SETUP 2 Stand For a good electrodes polarization a delayed startup every time t...

Страница 11: ...tact closed and N C contact open OFF is the status of N O contact open and N C contact closed ERROR MESSAGES If the display shows an error message use the following table ERROR 1 Buffer solution error...

Страница 12: ...on center for the recycling of electronic and electrical equipment For more information visit the online site When dismantling a pump please separate material types and send them according to local re...

Отзывы: