16
Calibrate the probe monthly or more if the application need higher precision.
The sensor need to be calibrated to the instrument it is connected to.
Two points calibration: 0 and a value close to working point.
For better results:
- connect the probe to the instrument;
- calibrate at plant temperature.
0 Calibration
Wash and rinse the probe. Calibrate in air.
2nd point calibration
Use a buffer solution with a value close to working point. Completely dip the probe into the buffer
solution.
CAUTION
: Make sure that no bubbles are inside the cavity of the probe.
Perform calibration according to the procedure described in the instrument manual.
CALIBRATION