Read carefully!

ENGLISH Version

R1-11-05

This manual contains important safety informations
about installation and use of this equipment. Ignoring
this informations could result in injuries or damages.

It is strictly forbidden to use this equipment with
radioactive chemicals!

O P E R A T I N G   M A N U A L   F O R
“ D I N   D I G I TA L   C D ”   C O N T R O L L E R

Содержание DIN DIGITAL CD

Страница 1: ...tant safety informations about installation and use of this equipment Ignoring this informations could result in injuries or damages It is strictly forbidden to use this equipment with radioactive che...

Страница 2: ...authorized personnel only Always discharge the liquid end before servicing the instrument Empty and rinse the liquid end before work on a pump which has been used with hazardous or unknown chemicals A...

Страница 3: ...AL WIRINGS 5 VIEW ON MENU 6 SETTING SETPOINTS 1 SETPN 6 Out mA 7 CALIBRATION 2 CALIB 8 DELAY 3 Delay 9 RESERVED 4 Reserv 9 PASSWORD SETUP 1 New Pw 10 MANUAL WORKING MODE 3 Serv 11 ERROR MESSAGES 11 Wa...

Страница 4: ...selection Press RIGHT key for about 4 seconds to switch off the instrument the display shows OFF Repeat the sequence to return to the normal operating mode To reset the instrument restore default sett...

Страница 5: ...Connect shield to ground 9 GND SEPR CONFIGURATION The instrument is set to work without a proximity sensor If needed proximity sensor turns the instrument into stand by mode if there isn t flow Instru...

Страница 6: ...oll MAIN menu is configured as follows 1 Setup This menu is divided into 1 Setpn 2 Calib 3 Delay 2 Param This menu is divided into 1 New Pw 3 Serv service mode SETTING SETPOINTS 1 SETPN Out 1 and Out...

Страница 7: ...de substitute percentual value to ON and OFF It is not possible to enter mixed values ex 100 for SP1 and ON for SP2 In this case it will not possible save them In proportional mode the instrument does...

Страница 8: ...ment will show Comp 3 This is the temperature compensation coefficient Modify this value according with the table on the buffer solution chemical product bottle using UP and DOWN keys Press E to confi...

Страница 9: ...aches the setpoint values Default value is set to 0 Select DELAY from SETUP menu The instrument will show OUT1 10 Sec Press RIGHT key The instrument will show OUT2 5 Sec Delay time can be set from 0 n...

Страница 10: ...et Using UP or DOWN keys from Param menu choose 1 New Pw press E to confirm The display shows NEW PW 0 0 0 0 Use UP and DOWN keys to modify th first digit Use RIGHT key to go on the next digit Press E...

Страница 11: ...ON is the status of N O contact closed and N C contact open OFF is the status of N O contact open and N C contact closed ERROR MESSAGES If the display shows an error message use the following table ER...

Страница 12: ...on center for the recycling of electronic and electrical equipment For more information visit the online site When dismantling a pump please separate material types and send them according to local re...