EMC INDUSTRIAL GROUP LTD
Settings
MW61A_IM_ALL_SV6.09d_en
22/47
Update with a PC
Updating Procedure
1.
Copy the ModUpdater program (e.g. ModUpdaterv6.01r10.exe) onto a PC. The name contains
the version number of the software to be installed.
2.
Before starting, it may be advisable to record the settings in the unit if you wish to restore
these after updating.
3.
Connect the ModWeigh unit to an RS232 COM port of the computer. For this you must use
the MAC cable and MAD adaptor.
4.
Connect power to the ModWeigh unit, 24Vdc.
5.
Double click on ModUpdater program.
6.
The upgrade process should then begin. Following the instructions given, usually you can
simply press ENTER at each stage.
7.
Restore and/or check settings.
If this does not work, then turn the power off to the ModWeigh instrument and back on again as in-
structed by the ModUpdater program.
If you have any problems, you can contact your supplier.
SETTING UP GUIDE
This guide suggests how a system can be setup. Full details on each of the settings can be found in the
‘Setup’ sections following the guide.
Settings
To calibrate the system, you must access and adjust various settings in the unit. The settings are ac-
cessed using their
Quick Key
numbers.
Quick Keys
To Access A Setting
1.
Press the Q key to access the setup menu.
2.
Press the
Quick Key
number to go directly to the setting. The UP & DOWN keys can be used
to go between sequential numbers.
3.
To select another setting, repeat the above process.
4.
Press and hold the BACK key to exit and return to normal operation.
To Adjust A Setting
1.
Press the EDIT key.
2.
If editing is locked, a password will be requested. Key in the password of 111 and press
ENTER.
3.
Use the keypad and/or UP and DOWN arrow keys to increase or decrease the displayed set-
ting.
4.
Holding either key down will automatically continually increase or decrease the value.
5.
Press ENTER to accept the changes or BACK to revert to the previous setting.
Quick Keys Reference
The following table is a quick reference to common settings.
Step
Quick Keys
BASIC SETTINGS
p24
Change the measuring range
212x
INPUTS
p25
Set the digital input terminal functions
221x
p26
Calibrate with weights
2232x
INTERNAL SIGNALS
p28
Set up how zeroing operates
2312x
p29
Change the limit settings
235xx
OUTPUTS
p31
Change the 4-20mA output calibrations
241x
242x
Windows™ PC
MAD
MAC
ModWeigh unit