4 Perform a test
1
2
3
4
1 Generator
Select the desired Waveform and Ground Injection as coupling type.
2 Connection cable
The calibration is performed including a 2m MC-cable. Connect the cable to the correct output.
The correct output is always indicated by a red LED.
3 Measurement Equipment
Use a HV differential probe to measure the open circuit voltage. The current is measured with a
current probe with a short circuit at the end of the connection cable.
4 Oscilloscope
The oscilloscope must have a minimum bandwith of 500 MHz.
Page 23/45