![embrace No Code Скачать руководство пользователя страница 23](http://html1.mh-extra.com/html/embrace/no-code/no-code_manual_2399626023.webp)
Control Solution Testing
The purpose of the control solution testing is to make sure the Embrace
Meter and the Test Strip is working properly. You should perform control
solution testing when:
• Using the meter for the first time
• Using a new bottle of Embrace No Code Blood Glucose Test Strips
• You left the test strip bottle cap open for a while
• You dropped the meter
• You suspect your meter and test strips are not working properly
• The blood glucose test results do not reflect how you feel
• Practice the testing procedure
IMPORTANT:
• Use only the Embrace™ Glucose Control Solutions with the Em-
brace No Code Test Strips. Other brands of control solution can pro-
duce inaccurate results.
• Always check the expiration date. DO NOT use control solutions if
expired.
• DO NOT FREEZE. Store the control solutions at room temperature.
You will need:
√ Embrace™ Control Solution (Level 1 or Level 2)
√ Your Embrace Meter
√ A new Embrace No Code Blood Glucose Test Strip
Step 1:
Take out an Embrace Test Strip from the test
strip bottle and close the bottle immediately.
Insert the test strip to turn on the meter.
Embrace N0 Code
Control Solution Testing
23