background image

OPERATION

MANUAL

AN AWARD WINNING, AFFORDABLE INFANT WARMER

Содержание nest

Страница 1: ...OPERATION MANUAL AN AWARD WINNING AFFORDABLE INFANT WARMER...

Страница 2: ...ior notice or obligation Contact your Embrace Innovations representative for the most current information The contents of the document are provided as is Please check with Embrace Innovations on the m...

Страница 3: ...design features 11 AccuTemp Heater user interface 12 3 3 WARMPAK 13 WarmPak design features 13 WarmPak user interface 15 3 4 WARMPAK TEMPERATURE INDICATOR 16 WarmPak temperature indicator design featu...

Страница 4: ...BabyWrap 32 CHAPTER 6 CLEANING MAINTENANCE AND STORAGE 33 6 1 CLEANING 33 AccuTemp Heater 33 WarmPak 33 BabyWrap 34 6 2 MAINTENANCE STORAGE 35 AccuTemp Heater 35 WarmPak 35 BabyWrap 35 6 3 PERIODIC CA...

Страница 5: ...m Customer Helpline toll free 1800 1033 123 ABOUT THE PRODUCT INTENDED USE AND CONTRAINDICATIONS Embrace Nest is intended to provide warmth to clinically stable newborns weighing between 1 5 to 2 5 ki...

Страница 6: ...NOTE NOTE provides useful additional information regarding a function or procedure DECLARATION OF LANGUAGE TRANSLATION The Embrace Nest comes to you as a package complete with labels user manuals and...

Страница 7: ...minated should be replaced immediately Should such repair or replacement become necessary Embrace Innovations recommends that a telephone or written request for service advice be made to Embrace Innov...

Страница 8: ...est is designed for use by clinically trained caregivers as an infant warmer in the clinic or during transport The Embrace Nest is NOT intended to be used by untrained caregivers or outside of clinica...

Страница 9: ...perature range as recommended by the World Health Organization WHO for a heated water mattress2 1 duration of at least 4 hours when ambient temperature is 25 C or higher 2 WHO reference number WS 420...

Страница 10: ...erature is low in a clinical setting After C section deliveries On examination tables In maternity wards before discharge To enable mother to child bonding while providing warmth 2 2 TRANSPORT USAGE W...

Страница 11: ...ak and BabyWrap The AccuTemp Heater is used to heat the WarmPak to operating temperature defined below This heated WarmPak serves as the source of warmth in the Embrace Nest The BabyWrap serves as an...

Страница 12: ...re Ever ready technology ensures heated WarmPak always ready for use Simple one button operation Built in safety features It is engineered to operate with minimal maintenance and is robust enough to w...

Страница 13: ...gress of heating Once ready the AccuTemp Heater will beep two times once when ready and once a minute after being ready The green WARMPAK READY indicator will remain illuminated The AccuTemp Heater wi...

Страница 14: ...esign features The WarmPak is made of heat resistant polyurethane It has two compartments one containing PCM and the other containing water When the WarmPak is heated using the AccuTemp Heater the PCM...

Страница 15: ...heating the newborn Some other unique design features of the WarmPak are User friendly indicator displays WarmPak readiness Certified non toxic materials Convenient SIM card design for ease of orienta...

Страница 16: ...erature Indicator indicates to the user if the WarmPak can be used or if it needs to be reheated The SIM card shape also serves to help the user in correctly orienting the WarmPak when placing it into...

Страница 17: ...or has been positioned such that it will be visible to the user ONLY when the WarmPak is correctly inserted into the BabyWrap as seen below The WarmPak temperature indicator alerts the user to potenti...

Страница 18: ...sed TOO HOT STATE There is a third state that will not be seen if the WarmPak is used correctly as indicated with the AccuTemp Heater In the event that the heater malfunctions due to improper use or a...

Страница 19: ...e and easy to sanitize with a disinfecting cotton swab The WarmPak compartment is heavily insulated on the outside to minimize heat loss from the WarmPak The flaps that cover the newborn also are suff...

Страница 20: ...vation window provides a view of the newborn inside the BabyWrap The medical staff can monitor the newborn s breathing skin color and general movement through this window Adjustable Straps The adjusta...

Страница 21: ...fabric is torn or if the insulation is spilling out Always ensure that the BabyWrap is washed thoroughly between uses across different babies Always periodically sanitize the interior surface of the B...

Страница 22: ...k after 6 months from date of first use Monitor the vital signs such as body temperature skin color breathing of the newborn in the BabyWrap at regular intervals Remove the newborn when the temperatur...

Страница 23: ...mp Heater is switched on all three LEDs will light up and buzzer will beep for one second The Heater performs a self check routine on itself If the AccuTemp Heater finds that all internal systems are...

Страница 24: ...er does not turn on even after replacing the fuse the AccuTemp Heater will need to be repaired or replaced In case of such an event occurring the user is advised to switch off power unplug the AccuTem...

Страница 25: ...leave it under the sun Store the WarmPak in a dry place preferably at room temperature Do not refrigerate the WarmPak Always store the WarmPak flat on a clean flat surface Do not fold it or store it i...

Страница 26: ...shown here 4 Ensure that the temperature indicator is visible through the temperature indicator viewing window 5 Do not place more than one newborn in a BabyWrap simultaneously 6 Use newborn safe cle...

Страница 27: ...Temp Heater Plug in the device and switch on the main power supply Check that the Power ON indicator illuminates This ensures that the AccuTemp Heater is receiving power and functioning properly In ca...

Страница 28: ...the PCM layer touching the heating surface To assist proper positioning of the WarmPak note the two indicators The Embrace Innovations logo should be upright and the word Embrace should be readable Th...

Страница 29: ...registered correctly by checking to see if the HEATING CYCLE ON indicator has turned on off 9 After approximately 35 minutes the buzzer will beep once to indicate that the heating cycle is complete a...

Страница 30: ...eater 10 The WARMPAK READY indicator will remain illuminated green until the lid is lifted or the Start Stop button is pressed This indicates that the AccuTemp Heater is maintaining the WarmPak at the...

Страница 31: ...indicator bar is in the TOO HOT state DO NOT use the WarmPak 3 Insert the heated WarmPak in the WarmPak compartment of the BabyWrap Ensure that it lies flat inside the WarmPak compartment NOTE Ensure...

Страница 32: ...ach of the newborn Setting up the BabyWrap 1 After placing the heated WarmPak in the WarmPak compartment place the BabyWrap on a flat surface with the newborn compartment facing up 2 Place the newborn...

Страница 33: ...ast 4 hrs in normal operating conditions7 It is therefore imperative to keep monitoring the WarmPak temperature indicator every 30 minutes to ensure that the WarmPak is within the operating temperatur...

Страница 34: ...ns that you normally use for cleaning plastics and other surface that the newborn comes in contact with AccuTemp Heater Sanitize AccuTemp Heater frequently All exposed surfaces of the AccuTemp Heater...

Страница 35: ...each new use Before placing a newborn in the BabyWrap sanitize the BabyWrap with a damp cotton swab cloth dipped in one of the above mentioned cleaning solutions Sanitize BabyWrap thoroughly for prolo...

Страница 36: ...e conditions outside of the recommended storage and operating range Do not handle roughly or drop the unit Do not place heavy objects on the AccuTemp Heater Avoid subjecting the AccuTemp Heater to ext...

Страница 37: ...ated WarmPak is not within the recommended operating range i e if the AccuTemp Heater is over or under heating the WarmPak do not use the AccuTemp Heater Contact us using the information in the sectio...

Страница 38: ...formance at varying ambient temperatures Time taken to heat WarmPak to ready state in AccuTemp Heater WarmPak duration in OK range 10 C approximately 35 minutes 1 5 hours 25 C approximately 35 minutes...

Страница 39: ...to the section on MATERIAL SAFETY at the end of this chapter for material safety information WARMPAK TEMPERATURE INDICATOR Dimensions 40 mm x 60 mm Weight 2 gm Operating Conditions Temperature 10 to 3...

Страница 40: ...RONMENTAL HAZARDS None identified FIRST AID MEASURES Eye In case of contact immediately flush eyes with plenty of water for at least 15 minutes Get medical attention Skin Wash skin with soap and water...

Страница 41: ...soap solution BABYWRAP INNER LINING MATERIAL BACTERIAL RESISTANCE OK as per AATCC 6538 9642 SKIN IRRITATION ISO 10993 10 Certified Safe NEGATIVE CYTOTOXICITY ISO 10993 5 Certified Safe NEGATIVE SKIN S...

Страница 42: ...race Nest infant warmer not including accessories which is defective if a claim is made during above mentioned period The following conditions procedures limitations apply to the Manufacturer s obliga...

Страница 43: ...be used in a clinic hospital nursing home or other medical institution under the supervision of the medical personnel who have been trained on using the product and whose skill and judgement determine...

Страница 44: ...nctions at its optimum when the core body temperature is between 36 5 C and 37 5 C This core body temperature condition is referred to as Normothermia Normal Operating Conditions With reference to the...

Страница 45: ...s safe to keep the newborn inside the BabyWrap The WarmPak should not be used if the WarmPak temperature indicator indicates that the WarmPak is outside the operating temperature range Phase Change Ma...

Страница 46: ...Embrace Nest Operation Manual 0602 50001 002 Page 45...

Отзывы: