background image

 

MeshNet Wireless Dual Channel 4 Wire RTD Input 

MN-RTD4W-H 

EDS-2030 

 

 

 

 

 

 

 
 

© 2019 Embedded Data Systems, LLC.; 2019 Fortune Drive; Lawrenceburg, KY 40342; Phone/Fax 502-859-5490    1 / 10

 

EN-USERMAN MN-RTD4W-H  V1.0  10.09.2020 

http://www.EmbeddedDataSystems.com

 

 

FEATURES

 

 

Wireless communication over the widely accepted 2.4 GHz frequency band 

 

High power radio 1200 meters (10x Gen 1 MN-ENV Radio)  

 

Buffer stores up to 2048 readings with time stamps 

 

Highly accurate (±0.1°C) and precise 4 Wire RTD probes readings 

 

RTD fault detection 

 

Programmable user friendly name & transmission rate 

 

Globally unique 64-Bit EUI (Extended Unique Identifier) 

 

Supports repeaters enlarging the wireless network and providing additional 
transmission routes for the sensor data 

 

Transmission strength reported on all devices, allowing troubled areas to be 
identified and addressed proactively (e.g. adding Repeaters) 

 

Robust Communication protocol includes acknowledgements and retries 

 

Energy efficient design optimizes battery life 

 

Low battery buffer only mode 

 

Optional pole mounting kit 

 

Applications include monitoring laboratory equipment and cooking processes 

 

Complies with FCC, Industry Canada and CE Marking requirements 

DESCRIPTION 

The 

MeshNet Wireless Dual Channel 4 Wire RTD Input Gen 2

, MN-RTD4W-H, offers an innovative way to monitor 

up to two  4-Wire RTD temperature probes. The RTD Input provides highly accurate and precise readings from the 
RTD  probes. The product also contains a memory buffer capable of holding 2048  sets of sensor readings, a high 
power radio with 10x Gen 1 sensor’s transmission range, RTD fault detection, and screw-down terminals.  Together 
these features offer an efficient and flexible system for monitoring wide temperature ranges from -200 to 800°C (RTD 
dependent) the temperature at multiple indoor and outdoor locations. 

Part Numbers 

Description 

MOQ 

MN-RTD4W-H

 

Two 4-Wire RTD Connections with Battery Power

 

MN-RTD4W-H-PLUG

 

Two 4-Wire RTD Connections Plugin Power with Battery Power Backup

 

45 

 

 

 

SYSTEM DESCRIPTION 

The MeshNet RTD Input is part of the MeshNet Wireless Sensor Ecosystem, designed to wirelessly monitor and control 
real world environmental conditions. The system is completed with the Embedded Data Systems MeshNet Controller 
(MN-CTRL). The controller communicates directly with the MeshNet RTD Input  and presents all data via Ethernet. The 
MeshNet RTD Input  and Controller communicate with each other using a proprietary protocol. An optional system 
component is Embedded Data Systems' MeshNet Repeater (MN-RPT-H) which is used to extend the range of 
communication between the Controller and Sensor Nodes. 
 

Controller:

 Each wireless system contains a MeshNet Controller. The Controller interfaces the wireless system to a local 

area network (LAN) or the Internet via Ethernet. It manages, reads/writes to the wireless devices, presents the data in 
easy to use formats such as HTTP and XML, and optionally pushes the data on a timed  basis  and/or per reception of 
wireless sensor packet to a HTTP Post server. 
 

Repeater:

 Repeaters are used to extend the distance a Sensor  Node can be located from the Controller. Many Sensor 

Nodes can communicate with the Controller through a single Repeater. Each Controller can manage up to 7 Repeaters, 
and there can be 1 Repeater between a Controller and Sensor Node. 
 

 

 

Содержание MN-RTD4W-H

Страница 1: ...or s transmission range RTD fault detection and screw down terminals Together these features offer an efficient and flexible system for monitoring wide temperature ranges from 200 to 800 C RTD dependent the temperature at multiple indoor and outdoor locations Part Numbers Description MOQ MN RTD4W H Two 4 Wire RTD Connections with Battery Power 1 MN RTD4W H PLUG Two 4 Wire RTD Connections Plugin Po...

Страница 2: ...will flash rapidly indicating that the sensor is now in an active state 8 Close the enclosure making sure that the screw down terminal is on the opposite side of the cable grommet to improve wireless performance by keeping the cables away from the wireless antenna 9 Lightly pull the cable to remove excess cabling from within the enclosure 10 Take the 4 screws included in your hardware bag and secu...

Страница 3: ...er higher numbers equate to stronger signals Due to hardware limitations the highest number that can be displayed is 34 dBm However the actual signal strength may be much higher DeviceRSSI This indicates the strength of the radio signal at the sensor Due to hardware limitations the highest number that can be displayed is 34 dBm however the actual signal strength may be much higher Address This is ...

Страница 4: ...dicates the oldest unsent buffered packet in the buffer When the BufferHeadPointer and BufferTailPointer are the same the buffer is empty RTD1ValueCalibration RTD2ValueCalibration It contains the factory calibration for the input channels Factory Reset Web soft reset Writing anything to the FactoryReset location will return all parameters except for the UserName and the Network Key to factory defa...

Страница 5: ...r increments until the counter is reset MessagePacketRetries The number of times wireless packets had to be resent due to no acknowledgment being received The maximum is 65535 after which it no longer increments until the counter is reset MessageFailures When no acknowledgment is received the packet is resent After 4 retries the message has failed The counter increments when a failure has occurred...

Страница 6: ... EDS RTD Sensor UserName owd_EDS2030_Tag owd_EDS2030_Data Description RTD Sensor with high power transceiver and power management Name EDS2030 Name EUI 801F12FFFECB7C0C EUI ControllerRSSI Units dBm 41 ControllerRSSI Address 1 Address Repeater 0 Repeater TimeStamp 2020 05 05 12 19 31 TimeStamp RTD1Temperature 250 000 RTD1Temperature RTD1Ohms 0 000 RTD1Ohms RTD1Fault 1 RTD1Fault RTD1CalibrationValue...

Страница 7: ...n be maximized by reading sending the sensor data as infrequently as possible Battery life is highly variable on a number of parameters 1 How often data is read sent 2 Extreme temperatures reduce the useful life of the battery 3 If the sensor is sending data through a repeater battery life will be reduced because the sensor has to wait longer for the acknowledgment 4 If the sensor has to reconnect...

Страница 8: ...ces When the name and password is requested enter admin for the name and eds for the password A page similar to below will appear 5 Insert AA batteries into sensor Once this is done the sensor will appear 7 in the Unlinked Devices section 6 Click the unlinked sensor A window will pop up confirming you want to link the sensor Click OK The sensor will move to the Linked Devices section upon its next...

Страница 9: ...Temperature humidity external probe support 2K memory buffer high power radio buffer only mode EDS2042 Up to 3 waterproof external temperature probes 2K memory buffer high power radio buffer only mode EDS REPEATER DEVICE IDS ID Features EDS2101 MeshNet Repeater with high power radio SPECIFICATIONS PARAMETER MIN TYP MAX UNITS Operating Temperature Range 40 85 C Accuracy 200 to 800 with Input at 21 ...

Страница 10: ... COMPLIANCE STATEMENT This Class B digital apparatus complies with Canadian ICES 003 Cet appareil numérique de la classe B est conforme à la norme NMB 003 du Canada EUROPEAN COMMUNITY EC DIRECTIVES CONFORMITY APPLICATION OF COUNCIL DIRECTIVE 2004 108 EC Standard to which Conformity is Declared EN 61326 1 2006 Emmissions EN 61326 2 3 2006 Immunity EN 61000 3 2 2006 A1 2009 A2 2009 Harmonics EN 6100...

Отзывы: