emaux 8552 Скачать руководство пользователя страница 4

P3    LCD Display and KEYPAD

NOTE:
Remove batter y from instruments 
that  you  do  not  plan  to  use  for  a 
month  or  more.  Do  not  leave 
batter y in instrument.

3. LCD Display and KEYPAD

The  main  display  shows  the 

measured ORP reading.

Hold = Data hold

4. OPERATION

1.  Remove  the  probe  cover  from  meter  to  expose 

the  electrode  out.  It  is  normal  if  you  find  white 

cr ystals  are  present  on  the  cap  or  electrode 

assembly. 

For  pH  meters,  please  always  make  the  sponge  wet 

to keep the electrode in a good storage condition.

Electrode

Remove the cover from the 

bottom of meter to expose 

the 

electrode.

2. Dip the electrode into the test solution. 

Press "      " to power on the meter and 

stir it to get a stable reading.

3.  Press  ”          to  freeze  current  readings. 

The text "Hold" will appear on the LCD. 

Press  "          "  again  to  release  the  hold 

mode. 

"

This meter will shut off automatically 20 minutes of 
inactivity.  For  operating  longer  time,  you  can 
disable the sleep mode.
To  disable  the  auto  power  off :  Before  power  on, 
pressing “      “ + “       “ keys simultaneously until a 
"

n

" appeared on the screen and then release keys to 

return to normal mode.
NOTE:  The  disable  sleep  mode  will  be  invalid  after 
ever y power off.

P4    AUTO POWER OFF (SLEEP FUNCTION)

4. Turn off the meter by pressing 

button. 

“      “ 

5.  Once  finishing  the  measurement,  clean  the 

electrode,  put  back  the  cap,  store  the  meter 

under the 0~50  ̊C.

5. AUTO POWER OFF (SLEEP FUNCTION)

6. 

MAINTENANCE

Preparation: 

Before using, remove the cover, then soak electrode 

into distilled water and rinse, and then take out and 

make dr y. Now the electrode is ready for use.

NOTE: 
Do not rub sensing element forcibly

Test the electrode: 

(1)  Put  electrode  in  buffer  solution  of  pH  7.00  with 

saturated  quinhydrone.  Af ter  stirring,  mV 

reading (E1) should be 86±15mV.

(2) Rinse electrode with distilled water, then set it in 

b u f f e r  s o l u t i o n  of  p H 4 . 0 1  w i t h  s a t u r a te d 

quinhydrone.  After  stabilizing,  record  mV  meter 

reading  (E2).  The  difference  between  E1  and  E2 

should be 165mV.

Storage: 

(1)  Rinse  the  electrode  with  distilled  water  during 

the inter val of each use.

(

2

Keep  ORP  electrode  wet.  If  not  use  for  long 

period of time, it should be rinsed and stored in 
the  provided  soaking  bottle  which  is  filled  with 
soaking solution.