For more information concerning the installation of the BioProTT
TM
FlowMeasurement System and how to
connect the BioProTT
TM
Clamp-On SL to the evaluation device, refer to the user manual of the respective
device.
•
Check and make sure that the used connector fits and is actually compatible with the
measurement system.
•
The connector of the BioProTT
TM
Clamp-On SL must be dry when plugged in.
•
While installing the BioProTT
TM
Clamp-On SL, be sure not to impair the function of the
tubing system.
•
Make sure to check the plug and the pins for damage before connecting the
BioProTT
TM
Clamp-On SL.
•
Do not use the BioProTT
TM
Clamp-On SL if any part of it is damaged.
6.3 Attaching the BioProTT
TM
Clamp-On SL to the Tube
In order to be able to measure the flow inside the tube, the BioProTT
TM
Clamp-On SL must be clamped onto
the tube.
•
Before you clamp on the sensor, make sure that the tube is not deformed, damaged, or
dirty.
•
Ensure that the tube size and the tube material used corresponds with the size of the
BioProTT
TM
Clamp-On SL.
•
Air in the tube can lead to errors in the measurement or interrupt the ultrasound coupling.
•
Ensure that there is no dirt or residue in the flow channel.
→
If necessary, clean the BioProTT
TM
Clamp-On SL before inserting the tube into the flow
channel.
First, make sure that the arrow on the sensor lid is aligned with the flow direction (positive flow display).
Then, to clamp on the BioProTT
TM
Clamp-On SL, follow the steps illustrated below:
1.
Open the spring-loaded lock and swing back the lid.
2.
Insert the tube into the channel.
3.
Close the lid and secure it with the help of the spring-loaded lock.
For more information regarding the setup and handling of sensor and tube, refer to our TechNote Issue #1
"
How to Optimize the Accuracy in Your Flow Measurement Application
Figure 2: Attaching the Sensor
BioProTT™ Clamp-On SL
Copyright em-tec GmbH
Page 11
D162-704 BioProTT™ Clamp-On SL – User Manual– V4.0