4.4 Switching to manual mode
Manual mode can be used to switch through the individual steps of the CSi/d Autoloader.
Fig. 3: Manual mode
Number
Component
Description
1
Button for manual control
Used to switch through the individual steps in
manual mode
2
Electrical connection safety
circuit
If the power supply unit of the CSi/d Autoloader
is plugged in there, the safety circuit is bridged
To switch to manual mode the CSi analyzer must be turned off and connected to the CSi/d
Autoloader. Proceed as follows:
1.
Pull the power supply unit from the back of the CSi/d Autoloader.
2.
Keep the manual control button (
1
) pressed and plug the power supply unit
into the designated electrical connection (
2
) in the safety circuit (see Fig. 3:
3.
Keep the manual control button (
1
) pressed for a few seconds (see Fig. 3:
Manual mode has now been activated and the individual steps can be activated using the
manual control button (
1
).
Switching to manual mode bridges the safety circuit.
Remove the power supply unit to exit manual mode.