Installation
15
2.8 Temperature adjustment
The furnace temperature is PC-controlled.
The CS-580A is delivered with 1350°C set temperature saved in a software profile.
After switching on the instrument at position 2 and starting of the UNI-software, the
heating up to the set temperature is started automatically. (The cable connection
between instrument and PC is in the scope of supply).
The actual furnace temperature and the set temperature are visible in the window
Heater.
You can change the set temperature in the window Configuration.
Fig. 4: Window Configuration – Heater – Temperature
2.9 Preheating the furnace
NOTICE
For heating up, the mains power switch of the analyzer has to be at position 2 and
the PC has to be switched on. The configuration of the BIOS has to be done by the
customer as needed.
During heating up to the set temperature, the ceramic combustion tube is under
high thermic stress. This stress is increased with increasing speed of the heating
up process.
For maximum service life of the combustion tube, the furnace can be heated up
slowly with a heating rate of 10°C/min. In order to avoid long waiting time until the
furnace heats up, the heating up of the furnace can be started automatically long
before the work begins. See chapter
Please note that the automatic preheating can only be started when the analyzer
temperature is below 100°C.
NOTICE
The analyzer is delivered with a default heating rate of 10°C/min. Much higher
heating rates are possible; however with respect to the service life of the
combustion tube, higher rates should only be used when necessary.
2.9.1 Configuration of preheating
1.
Open the window
Default
and choose
Configuration
.
Fig. 5: window
Default
2.
In the chapter
Furnace
you find
Preheat
.
Содержание CS-580A
Страница 2: ...2 Copyright Copyright by Eltra GmbH Haan Retsch Allee 1 5 D 42781 Haan Federal Republic of Germany ...
Страница 3: ...3 ...
Страница 23: ...Analysis 23 Weight 350mg Set the minimum analysis time to 50s Set the comparator level to 20mV ...
Страница 38: ...Function description 38 Fig 14 Gas flow ...
Страница 48: ...Ordering numbers 48 6 6 TIC module Fig 17 TIC module 35543 9008 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 ...
Страница 57: ......
Страница 58: ......
Страница 59: ......
Страница 60: ...Copyright Copyright by Eltra GmbH Haan Retsch Allee 1 5 D 42781 Haan Federal Republic of Germany ...