Pre-installation guide
DANGER
D1.0005
Mortal danger from electric shock
Exposed power contacts - High Voltage
−
An electric shock can cause injuries in the form of burns and cardiac
arrhythmia, respiratory arrest or cardiac arrest.
•
Set the mains switch of the analyser to position 0 and pull the
mains plug of the mains socket.
WARNING
W1.0021
Fire hazard / Risk of burns
Hot parts (crucibles, reagents,.. ) can fall down
−
Ignition of tables, floors, or any other surface the hot part falls on
−
Ignition of clothes and any other material
•
Set up the analyser in a flame retardant environment. Pay special
attention to the table, the floor and any other surface being in the near
of the analyser
•
Make sure that the surrounding space of the analyser stays clear of any
flammable material
•
Always wear suitable clothing
•
Keep the work environment clear of all materials that could catch fire
WARNING
Acceleration of fire / oxygen rich atmosphere
High purity oxygen carrier gas
–
increased risk of fires
•
Since the oxygen related risks heavily depend on the installation
situation in the laboratory, the user must make an own risk
assessment, evaluating the necessary steps to minimize the risks
introduced by handling oxygen from the source through the analyser
to the exhaust system.
Following requirements apply, when installing the analyzer:
Carrier gas
Oxygen 99.95 % pure; 2 - 4bar (30 - 60psi)
Mains power supply
230VAC ±10%, 50/60Hz; 20A fuse
CEE-Plug 230V, 32A
Содержание CS-580A
Страница 1: ...1 Manual Carbon Sulfur Analyzer CS 580A HELIOS REV0004 04 19 Translation ...
Страница 2: ...2 Copyright Copyright by Eltra GmbH Retsch Allee 1 5 42781 Haan Germany ...
Страница 56: ...56 8 2 Gas flow system ...
Страница 63: ...63 9 3 Bottom outside view Part Nr Description 11035 Cooling fan 11492 Oxygen pressure regulator ...
Страница 67: ...67 9 6 TIC module 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 ...
Страница 70: ...Copyright Copyright by Eltra GmbH Retsch Allee 1 5 42781 Haan Germany ...