Adjustments
73
–
The flow sensor (1)
–
The flow regulating valve (V6)
–
The flow control board HF 42
The flow sensor measures the gas flow rate, and converts its flow rate into an electrical signal
(DC voltage).
The HF 42 board compares this flow value with the flow value set by P1.
The signal resulting from the comparator controls a pulse width modulator.
Its output can be seen on TP 8, TP 15 is GND.
The pulse width modulated signal controls a power transistor, which drives the flow regulating
valve.
The duty cycle of the output signal determines the average DC voltage on the valve. In normal
operation with a flow of about 180 l/h and with not depleted chemicals, the DC voltage across
the valve coil is around 10V. This voltage can be measured between TP3 and TP17. This
voltage on the valve has the value needed at any moment for keeping the flow constant, in
other words for keeping the sensor voltage equal to the value set with P1.
4.1.2 Flow controller for resistance furnace operation
The flow sensor is common for both, induction and resistance furnace operation.
The sensor output is connected to both:
To the input of the HF42 board controlling the regulating valve V6 in induction operation and to
the input of the pump control board PC-1.3 controlling the pump in resistance furnace operation.
See chapter
“
Wiring diagrams
” and “
Gas flow diagram
”.
1
Содержание CS-2000
Страница 2: ...2 Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...
Страница 6: ...Notes on service instructions 6 ...
Страница 72: ...Adjustments 72 Fig 57 Gas flow regulation The flow controller mainly consists of the following components ...
Страница 77: ...Adjustments 77 Fig 61 IRC 1 3 ...
Страница 107: ...Service 107 Fig 91 Tubes to squeeze ...
Страница 120: ...Miscellaneous 120 6 1 2 Left hand side Fig 106 Left hand side view ...
Страница 122: ...Miscellaneous 122 6 1 3 Right hand side Fig 107 Right hand side view ...
Страница 129: ...Miscellaneous 129 6 1 7 Infrared cell Fig 111 Infrared cell ...
Страница 139: ...Miscellaneous 139 Fig 119 Left hand side power wiring 6 2 3 Valve wiring ...
Страница 140: ...Miscellaneous 140 Fig 120 Valve wiring 6 2 4 Control signal wiring Fig 121 Control signal wiring ...
Страница 142: ...Miscellaneous 142 6 2 6 IR cell internal signals Fig 123 IR current supply new version ...
Страница 145: ...Miscellaneous 145 6 3 2 Gas flow for induction furnace when furnace is open ...
Страница 146: ...Miscellaneous 146 6 3 3 Gas flow for resistance furnace ...
Страница 147: ...Miscellaneous 147 6 3 4 Valve contolling ...
Страница 160: ...Copyright Copyright by Eltra GmbH Retsch Allee 1 5 D 42781 Haan Germany ...