Maintenance
77
Fig. 50: Gas leaks check-furnace
•
Release the button (43), close the furnace, squeeze the tube (6) tight and press and hold
the button (43).
•
If the pressure on gauge (12) remains constant, then the furnace has to be checked for
leaks.
•
If the pressure shown on gauge (12) drops, then the leakage must be somewhere before
the furnace, between oxygen inlet and furnace.
6.14.1 Leaks in the furnace inlet system
•
After following the above instructions, check the inlet tubes (6) for leakage
6.14.2 Leaks in the furnace
•
After following above instructions, close the furnace, squeeze tight the tube (9), press and
hold the button (43), observe the pressure gauge (12).
•
If the pressure drops, then the furnace is leaking. Check whether the O-rings (16), (17)
and (18) are dirty or defective. See chapter
•
Check whether the combustion tube (15) is broken or cracked. See chapter
•
If the pressure remains constant, then the leakage must be after the furnace.
6.14.3 Leaks in the furnace outlet system
After following the instructions in the section
, check if the handle (8) is
properly shut, or else there will be a major gas leakage from the dust filter. Check the dust trap
(10) and the glass tube (11) for leakage.
Содержание CS-2000
Страница 2: ...2 Copyright Copyright by Eltra GmbH Retsch Allee 1 5 42781 Haan Germany ...
Страница 5: ...5 ...
Страница 63: ...Maintenance 63 Fig 35 Induction furnace Cone ...
Страница 80: ...Function description 80 Fig 51 Gas flow system ...
Страница 91: ...Miscellaneous 91 8 1 6 Furnace Fig 60 Furnace 71031 71010 ...
Страница 98: ...Miscellaneous 98 8 1 12 Tic module Fig 66 TIC module 35543 9008 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 ...
Страница 110: ...Copyright Copyright by Eltra GmbH Retsch Allee 1 5 42781 Haan Germany ...