Installation
20
When the analyzer’s mains power switch is set to position 2, a valve opens and the oxygen can
flow through the gas flow system. The flow rate is stabilized within a few seconds to 180 l/h and
it is displayed on the lower flow meter.
4.6 Resistance furnace-Temperature adjustment
Caution
C3.0096
Eye injury
Hot combustion tube.
Eye damage.
•
Avoid looking directly into the hot combustion tube
.
For eye protection
use the supplied protective glass.
WARNUNG
W4.0023
High temperature / Burning of the skin
Hot combustion tube
•
When the analyzer (HTF, resistance furnace) is switched on again after
switching off, the adjusted set point is driven immediately. Please
check this!
Depending on the material, the necessary temperature of the resistance furnace may be
different. See chapter “Applications – resistance furnace”. The procedure of adjusting the
resistance furnace temperature is described below.
Fig. 8: Resistance furnace-Temperature adjustment
•
Set the mains power switch to position 2.
•
Set the switch (14) to "Set Point" position.
•
Adjust 1350 on the display (15) by turning the potentiometer (13).
•
Set the switch (14) to "Furnace temperature" position.
•
The display (15) shows now the actual furnace temperature.
Содержание CS-2000
Страница 2: ...2 Copyright Copyright by Eltra GmbH Retsch Allee 1 5 42781 Haan Germany ...
Страница 5: ...5 ...
Страница 63: ...Maintenance 63 Fig 35 Induction furnace Cone ...
Страница 80: ...Function description 80 Fig 51 Gas flow system ...
Страница 91: ...Miscellaneous 91 8 1 6 Furnace Fig 60 Furnace 71031 71010 ...
Страница 98: ...Miscellaneous 98 8 1 12 Tic module Fig 66 TIC module 35543 9008 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 ...
Страница 110: ...Copyright Copyright by Eltra GmbH Retsch Allee 1 5 42781 Haan Germany ...