The top line of the display shows the selected measuring distance in cm, 
the bottom line shows the measured charge in Volts. If 999 V are reached 
during measurement, the unit automatically switches to the kV display 
range. The measured value is always displayed with three digits, 578 V; 
3,85 kV; 24,0 kV or 274 kV.
If the message "overflow!" appears on the display, select a larger measu-
ring distance!

Changing the measuring distance

Keep the "function/on" key pressed until the bottom line of the display 
shows "change distance" (approx. 2 sec). The topline shows the current 
distance in cm. You can now change the distance by briefly pressing the 
"function/on". The setting changes every time you press the key:
2 cm => 5 cm => 10 cm =>20 cm => 1 cm => 2 cm and so on. 
As soon as the desired setting appears in the display, wait 

without

 pres-

sing the "function/on" key until the bottom line again shows the current 
measuring value. The top line now shows the newly selected distance. 
This measuring distance is now used for any subsequent measurement.
The measuring distance results from the distance between the measuring 
electrode (behind the modulator wheel) and the measured object. The 
measuring electrode is located 3 mm ahead of the front of the unit. For 
this reason 3 mm must be added to every distance value when taking 
measurements from the front (aluminium plate).

Caution!

After switching the unit off and back on, the unit always returns to the 2 cm 
measuring distance setting!

HOLD

The instrument is equipped with a Hold function to "freeze" the measured 
value. The value measured at this point in time is held in the display by 
briefly pressing the "function/on" key once. The top line will show "Hold", 
the bottom line will show the measured value. The Hold function is swit-
ched off by again pressing "function/on" key briefly. The current measured 
value appears.

Содержание EM03

Страница 1: ...Operating Instructions Electric Field Meter EM03 BA en 1020 1804 F00029y ...

Страница 2: ...2 BA en 1020 1804_EM03 ...

Страница 3: ... 2 3 Proper Use 6 2 4 Work and operational safety 7 3 Operation 8 3 1 Startup 8 3 2 Taking measurements 8 3 3 Power supply 10 3 4 Battery monitoring 10 4 Maintenance 11 4 1 Changing the batteries 11 4 2 Zero setting 11 4 3 Calibration 11 5 Warranty 12 6 Technical specifications 13 7 Accesories and spare parts 13 Declaration of Conformity 15 ...

Страница 4: ... can be measured if 1 000 Volts are detected the instrument automatically switches over into the kV display range The electric field meter is a parametric amplifier The charges influenced by the electrical field generate an alternating current proportional to the field strength This alternating current is measured via a selective amplifier without drawing time averaged energy from the electrical f...

Страница 5: ...ring electrode is built into an aluminium front plate A modulation wind mill type wheel of the same star shape connected to frame rotates a short distance ahead of the electrode A ring electrode system enclosing the influence electrode serves as mechanical protection Fig 1 Outline of appli ance EM03 1 Spacer 2 cm 2 Modulator system measuring sensor with protective cap 3 Zero trimmer side mounted 4...

Страница 6: ...in the operating instructions refers to operations which if carried out improperly may result in damage to property 2 2 Technical advance The manufacturer reserves the right to make changes to the technical specifications without prior notice in order to adapt the units to state of the art engineering Eltex will provide the latest information on any changes or modifications in the operating instru...

Страница 7: ...ields 1 Hz cannot be measured with this instrument When measuring high static charges the electric field meter must be grounded using the alligator clip supplied with the unit The first measurement must be made from a sufficient distance to ensure that the maximum applied surface potentials are determined from a safe measuring distance Spark discharges onto the modulator system must be avoided ...

Страница 8: ... Fig 1 Pressing the key twice in brief succession during normal opera tion will switch the unit off Note The unit will switch off automatically after five minutes to avoid battery exhaustion Switch the unit back on for subsequent measurements 3 2 Taking measurements Remove the protective cap from the modulator system The preset measuring distance after switching on is 2 cm If this setting is used ...

Страница 9: ...out pres sing the function on key until the bottom line again shows the current measuring value The top line now shows the newly selected distance This measuring distance is now used for any subsequent measurement The measuring distance results from the distance between the measuring electrode behind the modulator wheel and the measured object The measuring electrode is located 3 mm ahead of the f...

Страница 10: ...s case the 9 V block battery must be replaced or recharged If the battery voltage drops below 7 0 V the bottom line of the display signals auto off for about 2 seconds The unit will then switch off automatically to avoid battery exhaustion and battery leakage Caution Use only alkaline or lithium 9 V block batteries or Ni Cd or Ni MH rechargeable batteries When using rechargeable batteries recharge...

Страница 11: ...nt Remove the battery and pull off the battery clip Push the battery clip over the new battery put the assembly back in the compartment and close the compartment 4 2 Zero setting Before each use of the unit we recommend a check of the zero point If with the modulator system protected e g by covering the modulator sys tem by hand and the testing person being grounded should it any time fail to disp...

Страница 12: ... the operating and assembly instructions spe cified by Eltex have been observed The warranty period begins on the date of delivery In the event of defects occurring during the warranty period the units or defective components will be repaired at Eltex Defective components will be replaced and installed free of charge If repairs are required at the customer s premises the costs for sending a techni...

Страница 13: ...Dimensions 114 x 68 x 25 mm L x W x H 141 mm with spacer and grounding socket Weight 175 g with battery Sensor star shaped influence measuring electrode Display 2 line alphanumerical LCD 2 x 12 digits Batteries 9 V alkaline or lithium block battery Type IEC 6F22 or Ni Cd or Ni MH batteries Operating period approx 10 hours continuous operation Measurement system calibration In the homogeneous field...

Страница 14: ...14 BA en 1020 1804_EM03 ...

Страница 15: ......

Страница 16: ... 67 69 D 79576 Weil am Rhein Phone 49 0 76 21 79 05 230 Fax 49 0 76 21 79 05 330 eMail static control eltex com Internet www eltex com Eltex offices and agencies The addresses of all Eltex agencies can be found on our website at www eltex com Z01007y ...

Отзывы: