EK280
Application Manual
138
9
Archive
i
Changes to the flexible archive structures which were programmed as factory settings may lead
to erroneous processing of read data in a subsequent system.
i
In case of e.g. a power failure, no archived data is lost as all archives are stored in non-volatile
memories.
9.1
Measurement period archive
In the measurement period archive meter readings and measured values are archived in regular measurement
periods
"MPer"
. The content of the measurement period archive can be shown on the display of the device via
the path
"Serv." > "Archives" > "Meas.period Archive"
.
i
The measurement period archive is a flexible archive.
Archive contents and the memory capacity can be adapted with corresponding settings via
"enSuite" parameterization software when the administrator lock is open.
Our applications team is available if you require further information.
9.1.1
Archive content
Pos. Address
Display
Description
Retrievable
1
1
01:01F8 GONo
Global Block number
S
2
03:0A20 ABNo
Archive Block number
A + S
3
01:0400 Time
Date and time
A + S
4
02:0300 Vb
Volume at base conditions
A + S
5
-
Vb
Meter progress Volume at base conditions
A
6
02:0302 VbT
Volume at base conditions total
A + S
7
-
VbT
Meter progress Volume at base conditions total
A
8
04:0300 Vm
Volume at measurement conditions
A + S
9
-
Vm
Meter progress Volume at measurement conditions
A
10
04:0302 VmT
Volume at measurement conditions total
2
A + S
11
-
VmT
Meter progress Volume at measurement cond. total
A
12
19:0161 p.MP
Pressure mean value of last meas period
A + S
13
15:0161 T.MP
Temp. interval mean value
A + S
14
23:0161 K.MP
Mean value K of last meas period
A + S
15
25:0161 C.MP
Mean value C of last meas period
A + S
16
02:0110 St.2
Instance-Status 2 (incl. Vn)
A + S
17
04:0110 St.4
Instance Status 4 (incl. Vb)
A + S
18
07:0110 St.7
Instance Status 7 (incl. p)
A + S
19
06:0110 St.6
Instance Status 6 (incl. T)
A + S
20
02:0100 StSy
System status
A + S
21
03:0A23 ErReg
Error register
A
22
03:0A22 Ev
Event causing data line to be archived
A + S
23
03:0A21 Check
Check sum result ( „CRC OK“ or „Error“)
A + S
The values can be retrieved in accordance with the information in the table via D = display and/or
I = interface.
When using an encoder then the original operating volume "Vo" (01 240) can also be set.