Connections to the EnCal 3000 Interconnection board
The drawing below shows the top lay-out of the Interconnection Board at the bottom of the unit. It contains all the con-
nectors for external cables (marked with grey). For flame retardant wiring use cables according an ISO norm. To fulfill
the FM regulations always use cables / wiring according to UL94 V-1 or equal for installation. All the wiring with the
other electronic boards is done in the factory, and may not be changed during field installation.
fig. 5-5 Interconnection Board: top view lay-out
J1
Processor board connection
J2
Channel 1 Connector
J3
Channel 2 Connector
J4
I/O Flat cable to Processor board
J5
TCP/IP connection between Inter-
connection board and Processor board
J6
Modbus RS485/232 connections
J7
Ethernet connection
J8
Solenoids terminal power suppl. 24 V DC
J9
Power supply out (12 VDC / 24 VDC)
J10
Analogue inputs (0-10 V DC)
J11
Solenoids terminal for valves of
stream selection (0-12 V DC)
J12
Power supply heater 1 (24 V DC)
J13
Power supply heater 2
(24 V DC)
J14
Dip Switch Power supply valves
(internal current supply circuit)
J15
Digital Inputs
(passive contact)
J17
Fan1 power supply connection (12 V DC)
J18
Fan2 power supply connection (12 V DC)
J19
Input
F106
Fuse analytical channels Module (6,3A)
F107
Fuse additional cabinet heaters (5A)
LD101
LED (internal 12 V- circuit)
LD102
LED (activation valves stream selection)
LK 301-304
4 jumper for selection between
RS232 and RS485
LK 305-307
3
jumper for selection of analogue input
(4-20 mA or 0-10 V)