background image

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

User manual 

 
 
 
 
 

 

 

Digital orbital shaker DOS-10M 

 
 

ELMI Ltd. 
21-133 Aizkraukles Str. 
Riga, LV-1006, Latvia  
T.: +371 67558743 
T.: +371 67551934 
E

[email protected] 

W: www.elmi-tech.com 

 

 

 

 
 
 

Содержание DOS-10M

Страница 1: ...User manual Digital orbital shaker DOS 10M ELMI Ltd 21 133 Aizkraukles Str Riga LV 1006 Latvia T 371 67558743 T 371 67551934 E info elmi tech com W www elmi tech com ...

Страница 2: ...ce 4 Control description 5 Using the device main menu 6 Orbital rotation mode 7 Auto reverse mode 7 Reciprocal mode 8 Vibration mode 8 Programming menu 9 10 Additional modes 11 Disinfection and cleaning 12 Transportation and storage 12 Warranty statements 12 Certificate of approval 13 Certificate of sale 13 ...

Страница 3: ...y dishes flasks Petri dishes etc Technical specifications Size of working platform mm 355х235 Speed of rotation RPM 10 500 Amplitude of platform movement mm 10 Maximum load kg 5 Working modes 4 Type of platform movement orbital Timer 1 m i n u t e 9 5 9 Environmental temperature C 4 up to 45 Relative moisture of atmosphere at 20 C temperature not more than 80 Voltage and frequency 110 220 V 50 60H...

Страница 4: ...vironments with aggressive or explosive chemical mixtures Do not operate the unit if it is faulty or been incorrectly installed For indoor use only Before using any cleaning or decontamination method except those recommended by the manufacturer check with the manufacturer that the proposed method will not damage the equipment Do not make modifications to the unit Biological safety It is the user s...

Страница 5: ...4 1 Network switch 2 Power plug 3 Working platform 4 Casing 5 Control panel 3 4 1 2 5 ...

Страница 6: ...ft right navigation keys and increasing decreasing parameter value keys Entering and exiting key in programming mode to set the step parameter Start stop key Setting key to manage the interface of device Entering and exiting key to programming mode ...

Страница 7: ... Reciprocal 4 Vibration To choose a mode go to the lowest level of the main menu and select the desired mode using the right or left navigation keys Adjusting the parameters of the mode Using the navigation keys select the desired mode and set required values for each parameter Press and hold the left or right navigation key to quickly change the value of a parameter ...

Страница 8: ...eters TIMER 1 minute 99 59 minutes continuous mode PRM 15 500 Auto reverse mode Auto reverse mode platform orbital motion mode with set and equal numbers of clockwise and counter clockwise turns Adjustable RPM time and the number of one way rotations Options TIMER 1 minute 99 59 minutes continuous mode PRM 15 RPM 500 RPM TURNS 5 25 ...

Страница 9: ... and operating time Options TIMER 1 minute 99 59 minutes continuos mode RPM 07 20 Angle 10 360 Vibration mode Vibration mode reciprocating type of movement of the platform imitating shake Ability to customize the speed settings time and amplitude of the shake up Options TIMER 1 minute 99 59 minutes continuos mode PRM 07 20 Amplitude 1mm 2mm ...

Страница 10: ...rogramming menu Ranges are from 1 to 99 or continuos mode can be used Create a custom cycle program by adding four steps and a number of program repeat cycles as required All settings are made using the navigation keys Enter and exit the programming step using ENTER key When setting the programming you can assign the desired operating mode 1 Orbital rotation mode 2 Auto reverse mode 3 Reciprocal m...

Страница 11: ...step menu s bottom line and select the pause mode set the time to 1 minute and press ENTER key 3 Select the 3rd step and press ENTER key 3 1 Go to the step menu s bottom line and select orbital rotation mode set the time to 2 minutes RPM to 200 and press ENTER 4 Select the 4th step and press ENTER key 4 1 Go to the step menu s bottom line and select the NO OPERATION mode and press ENTER 5 Go to th...

Страница 12: ...the PROGRAM key and switching the network switch off and on again To return to the advanced menu repeat the operation Interface adjustment menu In the interface menu you can adjust the display brightness sound volume and select the interface language All settings are made using the navigation keys To enter and exit the menu use the SETTINGS key English ...

Страница 13: ...an 80 and temperature range of 10 C up to 40 C It is not recommended to store the equipment more than 36 month Warranty statements Warranty applies to 24 month period from the date of purchasing Malfunctions that occur due to a fault of the manufacturer during warranty period are repaired free of charge Warranty is not valid in the following cases If the serial number label of the manufacturer is ...

Страница 14: ...ary for this class of device Quality control person name signature Date of manufacture Place for stamp Certificate of sale Organization Address Phone Vendor name signature Date of sale Manufacturers address 21 133 Aizkraukles Str Riga LV 1006 Tel 371 6755 8743 Tel 371 6755 1934 E mail info elmi tech com WWW www elmi tech com ...

Страница 15: ...14 Video instruction is available youtube com elmimedia videos ELMI Ltd 21 133 Aizkraukles Str Riga LV 1006 Latvia T 371 6755 8743 T 371 6755 1934 E info elmi tech com W www elmi tech com ...