ELMASLAR LIFETIME SA-PB 1 Скачать руководство пользователя страница 11

1.

 

Check whether all the hoses, container, secretion cap and bacteria filter are in perfect status. Defected or torn parts should 

be changed. 

 

2.

 

Check whether or not the hose connections are correct, and whether the secretion cap is inserted strongly and the container 

is inserted into the motor unit correctly. 

 

3.

 

Operate LIFETIME SA-PB device. After the device is operated, it starts to function in the lowest vacuum level and shows the 
current charge status. By visually checking the capacity indicator, control the charging status. If needed, charge the battery pack 
(see the chapter "5.4. Charging LIFETIME SA-PB"). 

 

4.

 

Close the suction piece of the front part with your thumb. 

 

5.

 

Operate the device and select -650 mmHg which is the maximum vacuum power.  LIFETIME SA-PB device should reach this 

 

vacuum power within approximately 10 seconds. 

It is understood that the device has reached this vacuum power by applying the following steps: 

 

All the LED lights from –100mmHg to –600mmHg become active stage by stage. 

-650 mmHg LED lights flash on and off. 

 

If the device reaches the set vacuum power for more than 10 seconds, it indicates that the suction power is low. Check for possible error 

(See the chapter "8. Faults and Troubleshooting". 

6.

 

Open the finger check hole of the end piece. 

7.

 

Close LIFETIME SA-PB device again. 

 

 

8.

 

Faults and Troubleshooting 

  

Description of Fault 

Underlying Reason 

Solution 

Device not operating, 

O/I

 indicator and 

 

 

capacity indicator band shows the device is 

Pump fault 

Enable the device to be repaired by the 

ready for operation. 

 

manufacturer or authorized specialist. 

 

Fuse fault 

 

 

Fuse fault in the vehicle socket 

 

 

Battery group over discharged. 

Repeat the charging and discharging 

Device not operating, 

O/I

 indicator shows the 

procedures for several times. If the result is 

 

device is ready for operation. 

 

the same, change the battery pack. 

 

Polar connections in the vehicle socket are 

Correct the incorrect polar connections and 

 

reverse. 

change the fuses if required. 

 

Connection between circle plate and battery 

Do the interconnection entirely. 

 

group is not well inserted. 

 

 

Device is not opened or closed. 

Electronic circle fault 

Make the device repaired. 

 

There are leakages in the suctioning parts of 

Check whether the hose connections, filter 

The device cannot reach the maximum 

cover and saliva cover be inserted 

the device. 

vacuum pressure of -650 mmHg for 10 

appropriately or not. 

 

seconds, but the capacity indicator band 

Battery pack is not sufficiently charged. 

Charge the battery pack. 

shows the device is ready for operation. 

Pump fault. 

Make the device repaired by the 

 

manufacturing plant or authorized specialist. 

 

 

At the -400 mmHg suction level or higher 

 

 

suction level LED light for -200 mmHg or -300  Bacteria filter is full. 

Insert a new filter. 

mmHg is continuously active. 

 

 

Charging procedure is not performed. 

External power feeding is so low. 

External power feeding should be between 
12.0 and 13.8V DC. 

 

 

 

Battery pack overheated. Charging is 

Make the temperature of the device lower 

 

impossible in the temperatures more than 

 

under +45°C. 

 

+45°C. 

 

 

 

Battery pack overcooled. Charging is 

Make the temperature of the device higher 

 

impossible in the temperatures less than 

Charging procedure is not performed. 

over +5°C. 

+5°C. 

100% LED light is not active when the 

 

Fuses fault. 

Make the device repaired by the 

charging procedure is completed. 

manufacturing plant or authorized specialist. 

 

 

 

Fuse in the vehicle socket fault. 

Change the fuse. 

 

Connection between circle plate and battery 

Do the connection thoroughly. 

 

group is not correct and entire. 

 

 

 

The charger does not have the required 

Use the original adapter/charger and/or 

 

features. 

accessories. 

 

The current supplied by the vehicle electric 

Check the vehicle electric installation. 

When the charging procedure is over 100% 

installation is not between 12.0-13.8 VDC 

 

LED light is not active. 

Capacity counter setting is abnormal. 

Discharge the device until it automatically 

 

stops and charge it until it is fully filled without 

 

 

 

 

Page 

11

 / 

14 

Содержание LIFETIME SA-PB 1

Страница 1: ...LIFETIME SA PB Suction Device Product Descriptions and Operating Manual ELMASLAR A MODELS LIFETIME SA PB 1 LIFETIME SA PB 2 LIFETIME SA PB 3 LIFETIME SA PB 4 Page 1 14...

Страница 2: ...n 8 5 3 Ending Aspiration 9 5 4 Charging LIFETIME SA PB 9 6 Hygienic Preparation 9 6 1 Preparation 9 6 2 Cleansing Disinfection and Sterilization 9 6 3 Connecting Pieces 10 7 Controlling Functions 10...

Страница 3: ...1 Overview Front view of LIFETIME SA PB device with reusable collecting container Back view of LIFETIME SA PB device Page 3 14...

Страница 4: ...Reusable Collection Container Accessories and spare parts Page 4 14...

Страница 5: ...visual warning for 6 seconds 2 Product Descriptions 2 1 Intended Use LIFETIME SA PB is a mobile portable electrically operated medical suction pump used temporarily in adults children and infants To...

Страница 6: ...ing with the multi use collection container the device must always stand upright because the ball of the overfill assembly cannot safely close the connection to the suction pump and secretions may ent...

Страница 7: ...template on the wall and align it with a water gage 3 Mark the holes for the assembly points 4 Make the holes by using a 3 mm sized drill bit 5 Screw the wall stabilizer by using the screws in the pac...

Страница 8: ...ve integrated into suction bag by the saliva the device should always be standing vertically If the fluid is scattered onto the overflow valve it becomes air resistant and the disposable aspiration ba...

Страница 9: ...cess ends all the lights on the indicator band will light up steadily Charging with wall stabilizer 1 Connect LIFETIME SA PB device to the wall stabilizer 2 Charging starts automatically a few seconds...

Страница 10: ...d is 20 minutes 6 3 Connecting Pieces Following the cleaning disinfection or sterilization procedures connect the pieces as follows 1 Screw the float of the overfill prevention device into the secreti...

Страница 11: ...are Correct the incorrect polar connections and reverse change the fuses if required Connection between circle plate and battery Do the interconnection entirely group is not well inserted Device is no...

Страница 12: ...ck is full but the charge status of the battery pack will not be displayed unless calibration is performed 12 Do a functional check see the chapter 7 2 Function Controlling 8 1 2 Calibrating capacity...

Страница 13: ...vacuum pressure in 12 V 650mmHg in sea level Aspiration hose 7 12 mm Length 1500 mm reusable container Bacteria filter Filter component of 1 m Motor power 55 W Rated voltage 12 V Maximum power 8 A Whi...

Страница 14: ...ransport damage caused by improper packaging when the device is returned Not performing the required or stipulated maintenance procedures Normal erosion due to operating The following parts are under...

Отзывы: