Maintenance
36
BA_ES xtra ST_EN_Vers.08.2016
© Elma Schmidbauer GmbH
9.4
Malfunctions
The following malfunctions are displayed as error messages on
the LED display:
Malfunction
Malfunction indication
Remedy
Bath temperature > 90°C
All LEDs of the temperature
indicator are flashing
Let the cleaning liquid cool
down to a temperature below
80°C; if necessary take out
some of the liquid and replace
by cold liquid.
As soon as the temperature is
below 80°C the ultrasound can
be restarted.
Error of temperature sensor
The LEDs of the temperature
indicator light one after the
other
Switch the device off and on
again. If the malfunction
persists, the device is defective.
Check the connection cable
between the electronic unit and
the temperature sensor.
If OK there is an error in the
electronic unit > replace the
electronic unit.
Communication error with
generator
The LEDs of the indicator for
the ultrasonic cleaning duration
light one after the other
Switch the device off and on
again. If the malfunction
persists there is an error in the
electronic unit > replace the
electronic unit
If one of these errors occurs, all buttons, except the On/Off
button, are deactivated.
Heating and the ultrasonics are switched off in each of the error
conditions.