Specifications
Temperature range D: S
-50 to 1000 oC(-58
℉
~ 1832
℉
) 30:1
Display resolution 0.1 oC(0.1
℉
) <1000
1
℉
>1000
Accuracy for targets:
Assumes ambient operating temperature of 23 to 25 oC (73 to 77
℉
)
-50 ~ 20
℃
(-58
℉
~ 68
℉
) ±2.5
℃
(4.5
℉
)
20
℃
~300
℃
(68
℉
~572
℉
) ±1.0% ±1.0
℃
(1.8
℉
)
300
℃
~1000
℃
(572
℉
~ 1832
℉
) ±1.5%
Repeatability
-50~20
℃
(-31~68
℉
) : ±1.3
℃
(2.3
℉
)
20~1000
℃
(68~1832
℉
): ±0.5% or ±0.5
℃
(0.9
℉
)
Response time 150ms
Spectral response 8~14um
Emissivity Digitally adjustable from 0.10 to 1.0
Over range indication LCD will show “----”
Polarity Automatic (no indication for positive polarity);
Minus (-) sign for negative polarity
Diode laser output <1mW,Wavelength 630~670nm,
Class 2 laser product
Operating temp. 0 to 50
℃
(32 to 122
℉
)
Storage temp. –10 to 60
℃
(14 to 140
℉
)
Relative humidity 10%~90%RH operating,
<80%RH storage
Power supply 9V battery, NEDA 1604A or IEC 6LR61,
or equivalent
Safety “ CE ” Comply with EMC
Note:
Field of View: Make sure that the target is larger than the unit’s spot size. The smaller the
target, the closer you should be to it. When accuracy is critical, make sure the target is at
least twice as large as the spot size.