background image

 

Sagab by Elma 911 

25 

 
 
SYMBOLS AND ANNUNCIATORS 
 

 

Continuity 

BAT 

   

Low Battery 

 

   

Diode 

DATA HOLD 

Data Hold 

AUTO 

   

AutoRanging 

AC 

   

Alternating Current or Voltage 

DC 

   

Direct Current or Voltage 

 

Specifications 

The instrument complies with: EN61010-1. 
Insulation: Class2, Double insulation. 
Overvoltage category: CATIII 600V, CATII 1000V. 
Display: 2000 counts LCD display with function indication. 
Polarity: Automatic, (-) negative polarity indication. 
Overrange: “OL” mark indication. 
Low battery indication: The “BAT” is displayed when the battery voltage drops below the 
operating level. 
Measurement rate: 2 times per second, nominal. 
Auto power off: Meter automatically shuts down after approx. 15 minutes of inactivity. 

Operating environment: 0 oC to 50 oC (32 oF to 122 oF) at < 70 % relative humidity. 

Storage temperature: -20 oC to 60 oC (-4 oF to 140 oF) at < 80 % relative humidity. 
For inside use, max height: 2000m 
Pollution degree: 2 
Power:
 One 9V battery , NEDA 1604, IEC 6F22. 
Dimensions: 150 (H) x 70 (W) x 48 (D) mm 
Weight: Approx.: 255g. 

Accuracy is given at 18 oC to 28 oC (65 oF to 83 oF), less than 70 % RH 
 
DC Voltage
 (Auto-ranging) 
Range 

Resolution 

Accuracy 

200.0mV 

0.1mV 

+0.5% of rdg + 2 dgts 

2.000V 

1mV 

 
+1.2% of rdg + 2 dgts 

20.00V 

10mV 

200.0V 

100mV 

600V(CATIII) 

1V 
 

+1.5% of rdg + 2 dgts 

1000V(CATII) 

+1.5% of rdg + 4 dgts 

Input Impedance: 10M

Ω

Maximum Input: 600V dc/ac rms.(CATIII) 
Maximum Input: 1000V dc/ac rms.(CATII) 
 

Содержание SAGAB Elma 911

Страница 1: ...Sagab by Elma 911 Multimeter Dansk Norsk 3 Svensk 13 English 23 EAN 5706445840274...

Страница 2: ...ftning af batteri 12 Udskiftning af sikringer 12 Svensk manual 13 S kerhetsinformation 13 Specifikationer 15 Anv ndning 18 Udskiftning af batteri 22 Udskiftning af sikringer 22 English User Manual 23...

Страница 3: ...rer o l som kan have jordpotentiale Hold din krop isoleret fra jordpotentiale ved at benytte t rt t j isolerende sko gummim tter eller andet godkendt isolerende materiale Kredsl b skal frakobles deres...

Страница 4: ...hvis potentiale i forhold til jord overskrider v rdien i dette eksempel 500V Dette symbol placeret ved en terminal indikerer at farlige sp ndinger kan v re til stede ved normalt brug For maksimal sik...

Страница 5: ...Instrumentet slukker automatisk efter ca 15 minutter uden brug Milj ved brug 0 C til 50 C 32 F til 122 F ved 70 relativ luftfugtighed Milj ved opbevaring 20 C til 60 C 4 F til 140 F ved 80 relativ lu...

Страница 6: ...telse 0 2A 250V og 10A 250V sikring Maksimalt Input 200mA dc eller 200mA ac rms ved uA mA skala 10A dc eller ac rms ved 10A skala AC str m Automatisk skala for A and mA Skala Opl sning N jagtighed 200...

Страница 7: ...s 2 000k 1 1 0 rdg 2 dgts 20 00k 10 1 2 rdg 2 dgts 200 0k 100 2 000M 1k 20 00M 10k 2 0 rdg 3 dgts Overbelastningsbeskyttelse 500V dc eller 500V ac rms Temperatur Skala Opl sning N jagtighed 50 C 1000...

Страница 8: ...sk v re i AutoRanging automatisk skala Dette v lger automatisk den optimale skala til den aktuelle m ling nskes manuelt skalavalg g res f lgende 1 Tryk p RANGE knappen I displayet slukkes AUTO 2 Tryk...

Страница 9: ...V AC positionen 2 Inds t den sorte testledning i den negative COM b sning og den r de testledning i den positive V b sning 3 Hold testproberne mod det m lte kredsl b 4 Afl s m lingen i displayet DC St...

Страница 10: ...kes foretaget 7 Tryk den sorte pr vespids mod den negative side af kredsl bet og den r de pr vespids mod den positive side af kredsl bet 8 Kredsl bet genindkobles 9 Afl s str mmen i displayet Str mtan...

Страница 11: ...testledning i den positive b sning 4 Tryk pr vespidserne mod dioden under test og noter afl sningen 5 Vend polariteten og noter afl sningen 6 Resultaterne kan tolkes som f lger A Hvis n m ling viser e...

Страница 12: ...NOTE Hvis instrumentet ikke fungerer korrekt kontroll r da sikringer og batteri for at sikre at de stadig fungerer korrekt Udskiftning af sikringer WARNING For at undg elektrisk st d skal instrumentet...

Страница 13: ...o dyl som kan ha jordpotential H ll din kropp isolerad fr n jordpotential genom att anv nda torra kl der isolerande skor gummimatta eller annat godk nt isolerande material Kretsar skall fr nkopplas fr...

Страница 14: ...ts med sp nning om potentialen i f rh llande till jord verskrider v rdet i detta exempel 500V Denna symbol placerad vid en anslutning indikerar att farliga sp nningar kan finnas vid normal anv ndning...

Страница 15: ...ing Instrumentet sl r av automatiskt efter ca 15 minuter utan anv Milj vid anv 0 C till 50 C 32 F till 122 F vid 70 relativ luftfuktighet Milj vid f rvaring 20 C till 60 C 4 F till 140 F vid 80 relati...

Страница 16: ...250V och 10A 250V s kring Maximal Input 200mA DC eller 200mA AC rms vid uA mA skala 10A DC eller AC rms vid 10A skala AC str m Automatisk skala f r A och mA Skala Uppl sning Noggrannhet 200 0 A 0 1 A...

Страница 17: ...et 200 0 0 1 1 2 rdg 4D 2 000k 1 1 0 rdg 2D 20 00k 10 1 2 rdg 2D 200 0k 100 2 000M 1k 20 00M 10k 2 0 rdg 3D verbelastningsskydd 500V DC eller AC rms Temperatur Skala Uppl sning Noggrannhet 50 C 1000 C...

Страница 18: ...automatisk skala Detta v ljer automatiskt den optimala skalan till den aktuella m tningen nskas manuellt val av skala g rs f ljande 4 Tryck p RANGE knappen I displayen slocknar AUTO 5 Tryck p RANGE kn...

Страница 19: ...sitionen 6 S tt den svarta testledningen i den negativa COM anslutningen och den r da testledningen i den positiva V anslutningen 7 H ll testprobarna mot kretsen som skall m tas 8 L s av m tningen i d...

Страница 20: ...retaget 16 Tryk den sorte pr vespids mod den negative side af kredsl bet og den r de pr vespids mod den positive side af kredsl bet 17 Kredsl bet genindkobles 18 Afl s str mmen i displayet Str mtangs...

Страница 21: ...n r da testledningen i den positiva anslutningen 4 Tryk pr vespidserne mod dioden under test og noter afl sningen 5 Vend polariteten og noter afl sningen 6 Resultaterne kan tolkes som f lger A Hvis n...

Страница 22: ...NOTE Hvis instrumentet ikke fungerer korrekt kontroll r da sikringer og batteri for at sikre at de stadig fungerer korrekt Udskiftning af sikringer WARNING For at undg elektrisk st d skal instrumentet...

Страница 23: ...d potential Keep your body isolated from ground by using dry clothing rubber shoes rubber mats or any approved insulating material Turn off power to the circuit under test before cutting unsoldering o...

Страница 24: ...ith respect to earth ground exceeds in this case 500 VAC or VDC This symbol adjacent to one or more terminals identifies them as being associated with ranges that may in normal use be subjected to par...

Страница 25: ...e 2 times per second nominal Auto power off Meter automatically shuts down after approx 15 minutes of inactivity Operating environment 0 oC to 50 oC 32 oF to 122 oF at 70 relative humidity Storage tem...

Страница 26: ...d Protection 0 2A 250V and 10A 250V Fuse Maximum Input 200mA dc or 200mA ac rms on uA mA ranges 10A dc or ac rms on 10A range AC Current Auto ranging for uA and mA Range Resolution Accuracy 200 0uA 0...

Страница 27: ...0 1 1 2 of rdg 4 dgts 2 000k 1 1 0 of rdg 2 dgts 20 00k 10 1 2 of rdg 2 dgts 200 0k 100 2 000M 1k 20 00M 10k 2 0 of rdg 3 dgts Input Protection 500V dc or 500V ac rms Temperature Range Resolution Accu...

Страница 28: ...meter is first turned on it automatically goes into AutoRanging This automatically selects the best range for the measurements being made and is generally the best mode for most measurements For meas...

Страница 29: ...meter 1 Set the function switch to the V AC position 2 Insert the black test lead banana plug into the negative COM jack and the red test lead banana plug into the positive V jack 3 Touch the test pro...

Страница 30: ...power to the circuit 9 Read the current in the display The display will indicate the proper decimal point value and symbol CLAMP ON ADAPTOR AC DC CURRENT MEASUREMENTS 1 Set the function switch to the...

Страница 31: ...on Note this reading 6 The diode or junction can be evaluated as follows A If one reading shows a value and the other reading shows OL the diode is good B If both readings show OL the device is open C...

Страница 32: ...e correct polarity 4 Put the battery door back in place Secure with the two screws WARNING To avoid electric shock do not operate the meter until the battery door is in place and fastened securely NOT...

Отзывы: