Elma 611B
side 15
Distance & Spot Size
As the distance (D) from the object increases, the spot size (S) of the area measured by the unit
becomes larger. The relationship between distance and spot size for each unit is listed below. The
focal point for each unit is 914mm (36”). The spot sizes indicate 90% encircled energy.
1.0in @
D:S= 30:1
D
S
30in.
25.4mm @ 762mm
2.0in @ 60in.
50.8mm @1524mm
Specifications
Temperature range
D:S
-50 ~ 1000
C (-58 ~ 1832
F)
30:1
Display resolution
0.1
C (0.1
F)
<1000
1
C (1
F)
>1000
Accuracy for targets:
Assumes ambient operating temperature of 23 ~ 25
C (73 ~ 77
F)
-50 ~ 20
C (-58 ~ 68
F)
±2.5
C (4.5
F)
20
C ~ 300
C (68 ~ 572
F)
±1.0% ±1.0
C (1.8
F)
300
C ~ 1000
C (572 ~ 1832
F)
±1.5%
Repeatability
-50 ~ 20
C (-31 ~ 68
F)
±1.3
C (2.3
F)
20 ~ 1000
C (68 ~ 1832
F)
±0.5% or ±0.5
C (0.9F)
Response time
150ms
Spectral response
8 ~ 14um
Emissivity digitally adjustable from 0.10 ~ 1.0
Over range indication
LCD will show “----”
Polarity:
Automatic (no indication for positive polarity);
Minus (-) sign for negative polarity
Diode laser:
Output <1mW,Wavelength 630 ~ 670nm,
Class 2 laser product
Operating temp.
0 ~ 50
C (32 ~ 122
F)
Storage temp.
–10 ~ 60
C (14 ~ 140
F)
Relative humidity
10%~90%RH operating,
<80%RH storage
Power supply
9V battery, NEDA 1604A or IEC 6LR61, or equivalent
Safety
“ CE ” Comply with EMC
Note:
Field of View: Make sure that the target is larger than the unit’s spot size. The smaller the target, the
closer you should be to it. When accuracy is critical, make sure the target is at least twice as large as
the spot size.
Содержание Elma 611B
Страница 1: ...MANUAL Elma 611B Dansk norsk 4 8 Svenska 9 13 English 14 19 EAN 5706445350001...
Страница 20: ...Ver 181113 hlo...