Elma 593R 

 

 

SPECIFICATIONS 

3.1  General specifications 

 

Measurement method

: Digital, triaxial measurement. 

 

Directional characteristic

: Isotropic, triaxial. 

 

Measurement range selection

: One continuous range. 

 

Display resolution:

 0.1mV/m, 0.1

A/m, 0.1

W/m

2

, 0.001

W/cm

2

 

 

Setting time

: Typically, 1s (0 to 90% of measurement value). 

 

Display refresh rate

: Typically, 0.5 seconds 

 

Display type

: Liquid-crystal (LCD), 4 digits. 

 

Audible alarm

: Buzzer. 

 

Units

: mV/m, V/m, 

A/m, mA/m, 

W/m

2

, mW/m2, W/m

2

W/cm

2

, mW/cm

2

 

 

Display value

: Instantaneous measured value, maximum value or maximum average value. 

 

Alarm function

: Adjustable threshold with ON/OFF. 

 

Calibration factor CAL

: Adjustable. 

 

Manual data memory and read storage

: 99 data sets. 

 

Rechargeable lithium-ion battery

: 3.6V, 1940mAh 

 

Battery charging time

: Approx. 3 hours 

 

Battery operation time

: Approx. 10 hours 

 

Auto power off

: 5 minutes. 

 

Operating temperature range

: 0

 to +50

  

 

Operating humidity range

: 25% to 75%RH 

 

Storage temperature range

: -10

 to +60

  

 

Storage humidity range

: 0% to 80%RH 

 

Dimensions

: Approx. 67(W)

60(T)

247(L)mm. 

 

Weight (including battery)

: Approx. 250g 

 

Accessories

: Instruction manual, Carrying case, AC adaptor (DC 5V 1.5A). 

 

3.2  Electrical specifications 

 Unless otherwise stated, the following specs hold under the following conditions: 

 The meter is located in the far field of a source, the sensor head is pointed towards the source. 

 Ambient temperature: +23

0

C     

 Relative air humidity: 25% to 75% 

 Sensor type: 

Electrical field (E)   

 Frequency range: 

10MHz to 8GHz

 

 Specified measurement range: 

 

CW signal (f > 10MHz)

: 20mV/m to 108.0V/m, 

53

A/m to 286.4mA/m, 

1

W/m

to 30.93W/m

2

0

W/cm

to 3.093mW/cm

2

 

 Dynamic range: 

Typically, 75dB   

 Absolute error at 1 V/m and 10 MHz: 

1.0dB 

 Frequency response: 

 

Sensor taking into account the typical CAL factor

:  

1.0dB (10MHz to 1.9GHz)  /    

2.4dB (1.9GHz to 8GHz) 

 

Isotropy deviation

: Typically, 

1.0dB (f>10MHz) 

 

Overload limit

: 10.61mW/cm

2

 (200V/m) 

 

Thermal response (0 to 

50

): 

0.5dB 

 

 

Содержание 593R

Страница 1: ...Manual Elma 593R ELECTROSMOG METER EN EAN 5706445340330...

Страница 2: ...5 4 OPERATION 6 4 1 Front panel controls description 6 4 2 LCD display description 8 4 3 Use E field sensor 9 4 4 Explanatory notes 9 4 5 Setting the meter 11 4 6 Making measurements 15 4 7 Manual da...

Страница 3: ...serve the operating instructions for equipment which is used to generate conduct or consumer electromagnetic energy Be aware that secondary radiators e g reflective objects such as a metallic fence ca...

Страница 4: ...magnetic flux density divided by the permeability of the medium Magnetic field strength is expressed in units of amperes per meter A m The meter uses of magnetic field strength for measurements only...

Страница 5: ...the range from 10MHz to 8GHz The non directional electric field and high sensitivity also allow measurements of electric field strength in TEM cells and absorber rooms The unit of measurement and the...

Страница 6: ...Operating temperature range 0 to 50 Operating humidity range 25 to 75 RH Storage temperature range 10 to 60 Storage humidity range 0 to 80 RH Dimensions Approx 67 W 60 T 247 L mm Weight including bat...

Страница 7: ...us Max instantaneous Max average 5 key Press this key to change unit s selector mV m or V m A m or mA m W m2 mW m2 or W m2 W cm2 or mW cm2 Press and hold this key while turning the meter on to disable...

Страница 8: ...device to the calibration factor setting mode press key to exit this mode 9 key Press this key one time to store one data set to memory Press and hold this key while turning the meter on to enter the...

Страница 9: ...he memory 9 Read manual data memory mode indication 10 Manual data memory address number 1 99 The manual data memory full indication 11 ALARM Alarm function on off or alarm setting indication 12 When...

Страница 10: ...ing to be moved in the 3 planes Simply point it at the target to make the measurement 4 4 Explanatory notes 4 4 1 Units of measurement The meter measures the electrical component of the field the defa...

Страница 11: ...ded for this application In many cases the measurement accuracy will be sufficient even if the frequency response of sensor calibration factor is ignored CAL can be set to 1 00 in such cases E Field t...

Страница 12: ...the units of measurement With the key as follows a Electric field strength V m b Computed magnetic field strength mA m c Computed power density mW m2 d Computed power density W cm2 X Y Z Z Y X 2 X Y Z...

Страница 13: ...Elma 593R 12 4 5 2 Setting the result mode Instantaneous result mode is automatically set when the meter is turned on With the key as followings MAX AVG MAX Instantaneous Max Instantaneous Max average...

Страница 14: ...turn on the meter to enter the alarm setting mode the mW m2 unit is flashing displayed 3 Press and keys to select the desired setting unit 4 Press key to enter the alarm value setting mode the one of...

Страница 15: ...tor setting mode the CAL SEt marks are displayed 3 Press or key to increase or decrease the value 4 Press key to store the new setting value and exit 4 5 5 Switching the alarm function on or off 1 Pre...

Страница 16: ...urned on the auto power off function is on 1 Press key to turn off the meter 2 Press and hold key and turn on the meter again to disable the auto power off function the symbol will disappear from the...

Страница 17: ...radiation Make measurements at those points where parts of your body are nearest to the source of radiation Note Use the Max average modes only when the instantaneous measurement values are fluctuati...

Страница 18: ...ss key again to exit 4 7 3 Deleting manual data memory measured values Once the memory is full you can clear the entire contents of the manual data memory 1 Press to turn off the meter 2 Press and hol...

Страница 19: ...cklight auto off 5 BATTERY CHARGING 1 Battery capacity Indication Battery Capacity Indication Battery Capacity 100 capacity 40 capacity 80 capacity 20 capacity 60 capacity 5 2 When the battery capacit...

Страница 20: ...v 210823 23 08 2021 14 34 31 Z Y X AVG MAX 2...

Отзывы: