AM852 Veterinary Anaesthesia machine User’ Manual
24
6.3Cleaning and sterilization of inspiration and expiratory unidirectional
valve
Loosen the cover of the inspiration and expiratory valves by rotating it anti-clockwise, then
clean valve cover, cover plate and valve mouth with the germfree gauze soaked in water
soluble disinfectant. After all parts cleaned and dried, recover the cover plate properly and
screw down the valve cover clockwise. Then user must check the airtightness and the
movement of the valve cover plate in accordance with the required regulation and checking
procedure. Please handle all parts with care lest valve cover and cover plate are bumped or
crushed.
6.4Cleaning and sterilization of absorber
Taking conditions of hospital, either suffocating sterilization (temperature no more than 50
℃
)or soak sterilization can be adopted in practice. When adopting soak sterilization, user
must dry all sterilized parts with high pressure air or oxygen after sterilization.
Warning:
Be sure to protect Valve carefully and prevent hard objects bumping them.
The damage of valve bodies (especially valve orifices) and valve flakes will
result in air leak so as to severe malfunction.