45
SECTION 3
CONTROLLING AND CUSTOMIZING STAINER FUNCTIONS
3.1 System Setup Menu
Enable Cyto Slide Tracking
Allows slide tracking in cytocentrifuge mode. See the Cytopro Rotor
Applications Manual (Aerospray Models 7xx2) (REF: RP-517) for complete
information.
Enable Manual Entry
If selected, allows manual entry of slide information using the keypad (limited
to 24 characters).
Enable Preventive Maintenance Tracking
To activate the tracking prompts for Preventive Maintenance Tracking, use the
following steps:
1 From System Setup, select
QC/Maintenance Tracking
.
2 Select
Enable Preventive Maintenance Tracking
.
3 Enter the information for the Daily, Weekly, and QC Slide prompts in
corresponding fields. See Using the Preventive Maintenance Log (Section
5.1).
Enable Reagent Tracking
To activate Reagent Tracking:
1 From the
QC/Maintenance Tracking
menu, select the reagent (A, B, C, D)
to be tracked.