Primary channel direct readout
store value measured directly by the analog-to-digital
converter for the primary channel. The value is presented for factory purposes and it is not
significant for user.
Auxiliary channel direct readout
store value measured directly by the analog-to-digital
converter for the auxiliary channel (not used for temperature measurement). The value is
presented for factory purposes and it is not significant for user.
Temperature
stores actual measured temperature. The value stored as unsigned integer with
range from 0 to 65535. 0 corresponds to the Operating range lower limit and 65535
corresponds to the Operating range upper limit. Temperature in the physical units can be
calculated using formula
t = Temperature / 65535 * (Operation range upper limit - Operation range lower limit) +
+ Operation range lower limit
where Temperature is the value read form the register 40082, Operation range lower limit is
the value read from registers 40008 – 40009 and Operation range upper limit is the value read
from registers 40010 – 40011.