PURELAB flex
1 & 2
- US
Operator Manual
Страница 1: ...PURELAB flex 1 2 US Operator Manual ...
Страница 2: ......
Страница 3: ... 7 6 Control of Substances Hazardous to Health COSHH 10 8 TECHNICAL SPECIFICATIONS 11 9 USEFUL CONTACT DETAILS 13 10 WARRANTY CONDITIONS OF SALE 13 Copyright Note The information contained in this document is the property of VWS UK Ltd trading as ELGA LabWater and is supplied without liability for errors or omissions No part of this document may be reproduced or used except as authorized by contra...
Страница 4: ...nd commission the PURELAB flex The Quick Reference Guide shows you how to perform day to day functions in order to get the most from your PURELAB flex without having to refer to the Operators Manual 1 5 Environment The PURELAB flex unit should be installed on a flat level surface in a clean dry environment The unit can also be wall mounted against a vertical wall capable of supporting the weight f...
Страница 5: ...ELGA PURELAB flex 1 2 US Operators Manual PURELAB flex 1 2 US MANU40562 Version 02 10 15 Page 3 2 YOUR GUIDE TO THE PURELAB FLEX ...
Страница 6: ...ilter Point of Use 90 Days 2 years LC197 Biofilter Point of Use 90 Days 2 years LC209 M2 Sanitization cartridge Single use 2 years Service Life is an estimate only and will depend on the application and feed water quality Please take care to ensure you order the correct consumable items 4 CUSTOMIZED OPERATION It s possible to customize your PURELAB flex by downloading additional features and chang...
Страница 7: ...ency decreases affecting water purity If either occurs you ll be prompted to change the UV Lamp WARNING UV C RADIATION IS HARMFUL TO THE EYES AND SKIN UV LAMP SHOULD ONLY BE OPERATED IN THE CHAMBER WARNING IT IS STRONGLY RECOMMENDED THAT DURING THE HANDLING OF THE UV LAMP CUT RESISTANT GLOVES ARE WORN MERCURY HAZARD DO NOT BREAK THE UV LAMP CONTAINS A SMALL AMOUNT OF MERCURY MAY CAUSE REDNESS OR I...
Страница 8: ...o the UV lamp carefully to ensure it doesn t break Step 3 Replace the UV lamp UNPACK new UV lamp CAUTION Take care not to touch the surface of the glass Ideally wear gloves handle with soft cloth and wipe the surface with alcohol before fitting into the housing SLIDE the new UV lamp into the UV housing CONNECT the lamp connector to the end of the UV making sure the connector is correctly aligned w...
Страница 9: ...er cable is connected to the mains electrical supply TURN OFF the water supply LOCATE the inlet filter assembly at the rear of the PURELAB flex Step 2 Dismantle the Inlet Filter UNSCREW the inlet filter REMOVE the mesh filter CHECK mesh filter for signs of damage replace or clean as necessary Step 3 Reassemble the Inlet Filter INSERT the mesh filter into its original position ensuring that it is f...
Страница 10: ...tched on Check the fuse in the power supply and replace if blown Unable to enter the configuration menu Press Auto Volume Dispense button to exit Auto Volume dispensing Purification Pack change reminder Reset Purification Pack change reminder Replace Purification Pack See Quick Reference Guide Sanitization reminder Reset Sanitization reminder Contact local distributor or technical support for furt...
Страница 11: ...designed to be submersed in water 7 3 Electricity The appliance coupler mains lead or power supply connected to the rear of the unit can be removed to isolate the power supply If access to this is restricted then it is recommended that access to the supply socket is easily available to disconnect the electrical supply WARNING ONLY USE THE APPLIANCE COUPLER MAINS LEAD AND POWER SUPPLY PROVIDED THE ...
Страница 12: ... Biostat WARNING DURING THE COMMISSIONING CYCLE TRACE RESIDUAL BIOSTAT IS FLUSHED FROM THE SYSTEM A MATERIAL SAFETY DATA SHEET IS AVAILABLE UPON REQUEST 7 6 Control of Substances Hazardous to Health COSHH Material safety data sheets covering the various replaceable Purification Packs are available upon request ...
Страница 13: ...iculates A 0 2 micron membrane pre filter is recommended for all non RO feeds to extend point of use filter life Iron Manganese 0 1 ppm Fe Mn Organics TOC N A N A 20 ppb TEMPERATURE 4 40ºC Recommended 15 25ºC FLOWRATE maximum requirement at 15ºC 2 l min Drain Requirements gravity fall with air gap None required 1 If feed water is in the range of 1 S cm 30 S cm you will need a pre conditioning cart...
Страница 14: ...y Dependant on feedwater Dependant on feedwater 5ppb2 Bacteria Typical when fitted with POU filter 1 CFU ml 1 CFU ml 1 CFU ml Bacteria Typical when fitted with Biofilter 1 CFU 10ml 1 CFU 10ml 1 CFU 10ml Endotoxin when fitted with Biofilter N A 0 001 EU ml 0 001 EU ml DNase when fitted with Biofilter N A N A 20 pg ml RNase when fitted with Biofilter N A N A 0 002 ng ml The PURELAB flex can be upgra...
Страница 15: ...nty provided herein and the data specifications and descriptions of the VWS UK Ltd products appearing in VWS UK Ltd published catalogues and product literature may not be altered except by express written agreement signed by an officer of VWS UK Ltd Representations oral or written which are inconsistent with this warranty or such publications are not authorized and if given should not be relied up...
Страница 16: ...ove warranty period is included in the warranty thereafter labour cost shall be at the customer s expense The exclusive remedy provided herein shall not be deemed to have failed of its essential purpose so long as VWS UK Ltd is willing and able to repair or replace any nonconforming VWS UK Ltd system or component part VWS UK Ltd Ltd shall not be liable for consequential incidental special or any o...