The purified water is then passed through the UV (UV1)
chamber where it is exposed to intense UV radiation to
provide continuous bacterial control and to promote the
cleavage of organic molecules before passing through an
absolute 0.2µm bacterial and particulate filter. The
difference in pressure between PG3 and PG4 can be
used to monitor the condition of the filter.
•
Finally the purified water passes through a:
•
Temperature sensor (TS2).
•
Water quality sensor (QS2), which measures the
resistivity of the water.
•
Air is drawn into the reservoir when water is used. To
maintain water purity in the reservoir and water
recirculation/distribution system, it is necessary to filter
the air. Water quality will be maintained within the
recirculation system using a composite vent filter, which
removes air-borne contaminants such as volatile organic
compounds and carbon dioxide as well as air-borne
particles and bacteria.
•
The
MEDICA - R200 US
includes the ability to drain the
reservoir completely via V4 and sanitize the installation
semi-automatically via V8 to ensure that bacterial levels
within the installation are controlled.
3.3
Flow diagram for MEDICA - R200 US
MEDICA - R200 US flow diagram