CENTRA S200/R200 - US

 

Operator Manual

Содержание CENTRA R200 US

Страница 1: ...CENTRA S200 R200 US Operator Manual...

Страница 2: ......

Страница 3: ...t and all restrictions on reproduction and use apply to all media in which this information may be placed VWS UK Ltd pursue a policy of continual product improvement and reserve the right to alter wit...

Страница 4: ...S 13 4 1 Unpacking the CENTRA 13 4 2 Positioning the CENTRA 13 4 3 Connecting the CENTRA 15 5 CONTROLS 19 5 1 PASSkeys 20 5 2 System Preferences 20 5 3 Setting Up Replacement Timers Reminders 27 6 COM...

Страница 5: ...ion selection 42 10 2 RO sanitization Process off 42 10 3 RO sanitization Process on 44 10 4 Recirculation sanitization 45 11 TROUBLE SHOOTING 47 12 CONSUMABLES AND ACCESSORIES 49 13 KEY TO CONTROL PA...

Страница 6: ......

Страница 7: ...Hz CENTRA R200HFR US 230V ac 50Hz 60Hz Unless stated the products will be referred to in the following generic terms CENTRA S200 US CENTRA R200 US 1 2 Use of this Manual This manual contains full deta...

Страница 8: ...NTACT WITH WATER 2 1 Electricity It is essential that the electrical supply to the CENTRA is isolated before any items are changed or maintenance work performed The external isolator providing power t...

Страница 9: ...US CENTRA R200 US Version 01 08 14 Page 3 2 4 Control of Substances Hazardous to Health COSHH Material safety data sheets covering the various replaceable components which contain chemicals are avail...

Страница 10: ...has been specifically designed to produce distribute and maintain purified water for laboratory medical and industrial applications Purified water may be distributed around the installation under pres...

Страница 11: ...mum flush time is completed Water quality is below set alarm point 4 minutes permeate flush has occurred without the permeate quality passing the alarm set point This will result in an alarm condition...

Страница 12: ...ided by an absolute 0 2 m bacterial and particulate filter The difference in delta P between PG1 and PG3 can be used to monitor the condition of the filter Finally the purified water passes through a...

Страница 13: ...ELGA CENTRA S200 US CENTRA R200 US Operator Manual CENTRA S200 US CENTRA R200 US Version 01 08 14 Page 7 3 3 Flow diagrams CENTRA S200 US flow diagram CENTRA R200 US flow diagram...

Страница 14: ...treatment Calcium Ca ppm as CaCO3 250 None 250 Softener and or use very low RO recovery Total chlorine Cl ppm 0 1 None 0 1 to 0 5 20 inch carbon 0 5 Cylinder of carbon sized correctly to obtain 0 1ppm...

Страница 15: ...Connections CENTRA S200 US CENTRA R200 US Inlet bsp bsp Sanitization Drain bsp bsp Reservoir drain bsp bsp Ion exchange cylinder inlet Install in Distribution Loop if required bsp Ion exchange cylind...

Страница 16: ...utput timed Selectable On Off Remote display configuration Mode address selection Sanitization reminder Selectable On Off Indicators Pre treated feedwater quality Conductivity Product water quality Re...

Страница 17: ...emote display Product Water Specification Voltage freq R S 200 R S 200 HFV R S 200 HFR Flowrate exit from unit l min USG min 230 50 115 60 16 4 2 18 4 8 30 8 38 10 30 8 38 10 Distribution Loop Pressur...

Страница 18: ...US Operator Manual ELGA Page 12 CENTRA S200 US CENTRA R200 US Version 01 08 14 3 5 RO Permeate Flowrate vs Temperature Nominal Permeate Flowrate vs Temperature 0 5 10 15 20 25 100 125 150 175 200 225...

Страница 19: ...ater treatment technologies are selected for your application 4 2 Positioning the CENTRA Before installation and operation of the CENTRA unit please read and observe the following points Environment T...

Страница 20: ...LATE THE INCOMING ELECTRICITY SUPPLY BEFORE REMOVING COVERS OR COMMENCING MAINTENANCE WORK COULD RESULT IN DEATH Pipework and Pipework Installation The precise details of each installation will vary a...

Страница 21: ...Hydraulic connections 1 LOCATE all the hydraulic connections 2 FOLLOW the manufacturer recommendations for the selected pipework systems and install suitable adapters into the connection manifolds of...

Страница 22: ...200 US CENTRA R200 US Operator Manual ELGA Page 16 CENTRA S200 US CENTRA R200 US Version 01 08 14 Fig 1 Schematic of typical CENTRA S200 US installation Fig 2 Schematic of typical CENTRA R200 US insta...

Страница 23: ...he cable through the rear of the unit to the power socket Note The power cable should be routed avoiding signal cables and sensitive circuitry 5 LOCATE the electrical cover on the top of the unit and...

Страница 24: ...int on the electrical back plate along with the earth wire from the electrical cover 15 REPLACE the earthing nut and washers and TIGHTEN WARNING THIS APPLIANCE MUST BE PROPERLY EARTHED ENSURE THE EART...

Страница 25: ...s Details of how to use the controls will be given in the appropriate sections The CENTRA control panel has a range of control icons General operational icons are as follows Further icons are describe...

Страница 26: ...trolled touch pad buttons that are used to control set up and process control functions When the CENTRA unit is started for the first time after installation the following steps should be carried out...

Страница 27: ...crease or DOWN to decrease minute 5 PRESS SHIFT to accept the minute setting and set the seconds to 00 6 PRESS TICK Step 3 Date Used to instigate change reminders it will appear on printed records 1 P...

Страница 28: ...ing This screen allows preferred displayed units of water purity to be set to either M cm or S cm This only affects the measurement in the recirculation loop 1 PRESS SCROLL to change mode M cm or S cm...

Страница 29: ...and do not change to S cm regardless of the water purity unit displayed setting OR TICK to proceed to step 11 2 PRESS TICK Step 11 RO permeate temperature alarm settings TS1 This screen is used for se...

Страница 30: ...for the installation can be selected During OFF periods the unit will automatically enter recirculation for a period of 10 minutes every two hours to maintain water purity within the distribution loo...

Страница 31: ...UTO MANUAL restart option If auto restart is selected the unit will automatically restart after a power failure In manual mode the unit will remain in standby 1 PRESS SCROLL to change mode ON OR PRESS...

Страница 32: ...ation Guide RS485 Local area Network design considerations and installation guide Available from ELGA LabWater Technical Support 3 POWER ON remote control stations 4 PRESENT user PASSkey any colour at...

Страница 33: ...POINT to load new sanitization PASSkey identification 5 REPEAT instruction 2 and 3 until all sanitization PASSkeys are registered maximum of 8 users 6 PRESS TICK to complete settings 5 3 Setting Up Re...

Страница 34: ...d OR PRESS CROSS to abort reset 3 PRESS TICK Step 5 0 2 m filter replacement date 1 PRESS RESET button to reset 0 2 m Filter Replacement Date OR PRESS TICK to accept Replacement Date and proceed to Sa...

Страница 35: ...splay will illuminate The initiation is complete when the display shows three lines of text 3 CONFIRM that the display states the correct unit type If incorrect contact your local ELGA LabWater repres...

Страница 36: ...irculation system will start automatically 4 ALLOW system to run and manually BLEED air from filters installed in the recirculation loop 5 CHECK for leaks around the installation Step 6 Installation o...

Страница 37: ...the distribution loop should be rated to 6 bar to ensure the equipment remains within safe operating limits To adjust the pressure in the distribution loop requires specialist equipment and it is reco...

Страница 38: ...NITIZE the system as described in Section 10 Sanitization Procedures Note It is good practice to sanitize new installations during commissioning Step 10 Fit 0 2 m in line filter LC160 or FILT50105 for...

Страница 39: ...5 minutes every two hours to maintain water purity around the distribution loop The sleep period will be entered once the following conditions have been met RO has filled reservoir to 350l or RO is of...

Страница 40: ...ressure at end of distribution loop Reservoir water volume Permeate quality Permeate temperature 8 3 Recirculation Displayed value Auto scroll Manual scroll Product quality Product temperature Flowrat...

Страница 41: ...lter Air filter should be replaced in the following circumstances Six months after initial installation Reservoir overflow indicated by incorrect reservoir volume Quality tests indicate replacement is...

Страница 42: ...CENTRA S200 US CENTRA R200 US Operator Manual ELGA Page 36 CENTRA S200 US CENTRA R200 US Version 01 08 14 14 RESET the consumable reminder as described in Section 5 3 Step 2...

Страница 43: ...venient point of use Note Some water may be spilt 3 OPEN front of unit by removing securing screws 4 LOCATE filter spanner found inside of front panel door 5 LOCATE the blue filter bottom left hand si...

Страница 44: ...wer 2 DEPRESSURIZE the distribution loop by opening a convenient point of use OR PRESSING the red knob on the top of the filter housing Note Some water may be spilt 3 OPEN front of unit by removing se...

Страница 45: ...raw lamp WARNING IT IS RECOMMENDED THAT DURING THE HANDLING OF THE LAMP CUT RESISTANT GLOVES BE WORN 5 HOLD lamp in gloved hand and PULL firmly on electrical connector to remove 6 DISCARD lamp CAUTION...

Страница 46: ...V6 See flow diagram 3 3 in Section 3 4 OPEN bypass valve V7 5 REMOVE connections to ion exchange cylinder making a note of the connection orientation 6 REMOVE cylinder and contact you local ELGA LabWa...

Страница 47: ...rs greater protection from chemical entering the recirculation loop Or Process on Method to allow sanitization of the RO whilst retaining the ability to supply the final application Refer to the corre...

Страница 48: ...f filter bowl 5 ATTACH chemical tube TUBE37548 to underside of filter housing WARNING ALWAYS ENSURE SAFETY GLASSES AND PROTECTIVE GLOVES ARE WORN WHEN HANDLING THE SANITANT 6 POUR 150ml of Minncare in...

Страница 49: ...mplete an alarm will sound 4 seconds 11 REMOVE filter bowl 12 Remove chemical tube TUBE37548 13 INSTALL new filter LC159 or refit existing filter 14 REFIT filter bowl tighten with filter spanner 15 PR...

Страница 50: ...RO particle filter bowl FH1 using filter spanner 6 Remove filter LC159 and discard if blocked or 6 months old 7 Empty remaining water out of filter bowl 8 Attach chemical tube TUBE37548 to underside...

Страница 51: ...PTY remaining water from bowl 8 WARNING ALWAYS ENSURE SAFETY GLASSES AND PROTECTIVE GLOVES ARE WORN WHEN HANDLING THE SANITANT 9 ATTACH chemical tube TUBE37548 to underside of the filter housing 10 PO...

Страница 52: ...stallation replacement of ion exchange cylinder R200 only 21 REPLACE LC160 FILT50105 0 2 m in line filter See section 9 3 Replacing 0 2 m in line filter 22 STORE sanitization tube TUBE37548 in safe pl...

Страница 53: ...ty value Mute alarm Check alarm set value is correct See Section 5 2 System preferences Check cylinder is in line and not isolated from the distribution loop Replace ion exchange cylinder See section...

Страница 54: ...and secure pipework to stop vibration Check air is not being drawn into the recirculation loop If problem persists call Customer Services PASSkey not recognized PASSkey not valid Request registration...

Страница 55: ...treatment filter 6 months 2 years LC160 0 2 m in line filter NOT HF 6 months 2 years CFYCZD1460 0 2 m filter 20 HFV and HFR 6 months 2 years Service Life is an estimate only and will depend on the app...

Страница 56: ...ey Auto restart Leak detection Scroll back Locked Step back Level sensor disconnect Bell Audible alarm Menu Standby Mute alarm Bypass DI Night Calibration point Node Cancel sample Option OFF Cancel Op...

Страница 57: ...Page 51 ICON DESCRIPTION ICON DESCRIPTION Fast timer Reset Reservoir feed Right Fill Sample Filter Sanitization PASSkey Insert filter RO Sanitization reminder Filter recirc loop Sanitization rinse Add...

Страница 58: ...ure icon flashing Inlet pressure to RO pump low check pre treatment filter are not blocked and water supply is on Ensure sufficient feed pressure Reservoir feed quality alarm measurement flashing RO o...

Страница 59: ...53 13 2 Replacement Timers Screen Replacement Reference Vent filter replacement See section 5 3 step 2 Pre treatment filter replacement See section 5 3 step 4 0 2 m in line filter replacement See sect...

Страница 60: ...omer notifies VWS UK Ltd promptly of any such defect The exclusive remedy provided herein shall not be deemed to have failed of its essential purpose so long as VWS UK Ltd is willing and able to repai...

Страница 61: ...stained by any customer from the use of its process systems Products or components manufactured by companies other than VWS UK Ltd or its affiliates Non VWS UK Ltd products are covered by the warranty...

Страница 62: ...TAILS ELGA LabWater Lane End Industrial Park High Wycombe Bucks HP14 3BY UK Tel 44 0 203 567 7300 Fax 44 0 203 567 7205 E mail techsupport elgalabwater com For the address of your nearest ELGA LabWate...