CENTRA-R 60/120 US Quick Start Manual 

 

CENTRA-R 60/120 US MANU40589 Version 02 - 02/16

 

Page 3 

1.  INTRODUCTION 

1.1  Use of this Manual 

ELGA LabWater strives to produce manuals that are as simple and accurate as possible.  However, should you 
feel they can be improved in any way please email us at 

[email protected]

. 

This manual guides you through the installation, commissioning and basic operation of the CENTRA-R 60/120 US 
allowing  you  to  obtain  a  guaranteed  supply  of  purified  water  to  meet  your  requirements.    For  more  detailed 
information refer to the Operator Manual supplied on CD. 

1.2  Customer Support 

Service  support  and  consumable  items  are  available  from  your  local  supplier  or  distributor.  Please  contact 

[email protected]

 for further information. 

1.3  Health and Safety Notes 

 

WARNING!

 

WARNINGS  ARE  GIVEN  WHERE  FAILING  TO  OBSERVE  THE  INSTRUCTIONS 
COULD RESULT IN INJURY OR FATALITY.

 

 

 

CAUTION!

 

Cautions are given where failure to observe the instructions could result 
in damage to the equipment, associated equipment and processes.

 

1.4  Electricity 

 

WARNING!

 

THIS APPLIANCE MUST BE EARTHED.

 

It  is  essential  that  the  electrical  supply  to  the  CENTRA  -  R  60/120  is  isolated  before  any  items  are  changed  or 
maintenance work performed. 

The external isolator providing power to the unit should be positioned so that it is easily accessible. 

1.5  Pressure 

The  water  supply  should  be  isolated  and  residual  pressure  released  prior  to  removal  of  any  components  or 
carrying out work on the unit. 

1.6  Sanitization Chemicals 

During the sanitization cycle EfferSan

™ Multi-purpose Disinfecting Tablets or Minncare Cold Sterilant is used and 

relevant  safety  guidance  is  included  in  this  handbook.  Please  refer  to the  manufacturer  for  material  safety  data 
sheets.  

 

EfferSan

™ and Minncare Cold Sterilant are EPA registered as a sterilant, high level disinfectant, and sanitizer.  

 

EfferSan

™  and Spent Minncare Cold Sterilant are acidic and requires normal neutralization as specified by your 

local state and local regulations. 

 

1% of Minncare Cold Sterilant has a pH of 3.5. 

1.7  Control of Substances Hazardous to Health (COSHH) 

Material  safety  data  sheets  covering  the  various  replaceable  components  are  available  upon  request.    Contact 
your local ELGA LabWater distributor. 

 

 

Содержание CENTRA-R120-US

Страница 1: ...CENTRA R 60 120 US Quick Start Manual ...

Страница 2: ......

Страница 3: ...on reproduction and use apply to all media in which this information may be placed VWS UK Ltd pursue a policy of continual product improvement and reserve the right to alter without notice the specification design price or conditions of supply of any product or service ELGA LabWater is a trading name of VWS UK Ltd ELGA is the global laboratory water brand name of Veolia Water ELGA PURELAB MEDICA a...

Страница 4: ...OSHH 3 2 INSTALLATION 4 2 1 Unpacking CENTRA R 60 120 US 4 2 2 External services and connections 4 2 3 Connecting CENTRA R 60 120 US 5 3 COMMISSIONING 6 4 CONSUMABLE INSTALLATION 7 4 1 Protek L1 or L2 Pre treatment cartridge 7 4 2 By pass block 8 4 3 System purge 8 5 COMMISSIONING SANITIZATION 9 5 1 Liquid Sanitization 9 5 1 1 RO Sanitization 9 5 1 2 Recirculation Sanitization 12 5 2 Tablet Saniti...

Страница 5: ...pment and processes 1 4 Electricity WARNING THIS APPLIANCE MUST BE EARTHED It is essential that the electrical supply to the CENTRA R 60 120 is isolated before any items are changed or maintenance work performed The external isolator providing power to the unit should be positioned so that it is easily accessible 1 5 Pressure The water supply should be isolated and residual pressure released prior...

Страница 6: ...taken into consideration Water Supply Local potable supply with a means of isolation 2 2 External services and connections Connection Termination Specification Unit of Measure Metric US Imperial Feed water inlet BSP female thread Flow 9 l min 2 4 USG min Pressure 2 to 6 bar 30 to 88 psi Temperature 4 to 40 C 39 2 to 104 F Pressurised drain RO BSP female thread Maximum flow 10 l min 2 7 USG min Max...

Страница 7: ...ccess to the sanitization port Drain hose connection 2 CONNECT drain to local drain with tube supplied Notes Ensure air gap is maintained from outlet of tube into a standpipe or surface drain The drains can be extended to a maximum length of 5 metres from the product using 15 mm PEXs rigid pipe It is strongly recommended that the hoses are secured to ensure they are always directed into the drain ...

Страница 8: ...tting clock time Step 2 Clock 1 PRESS flashing CLOCK clock icon 2 PRESS to edit time OR PRESS to proceed to Step 3 3 PRESS and HOLD to increase or to decrease hour 4 PRESS to shift cursor onto minute 5 PRESS to increase or to decrease minutes 6 PRESS 7 PRESS Set date screens Step 3 Date 1 PRESS to edit date OR PRESS to proceed 2 PRESS and HOLD to increase or to decrease day 3 PRESS to shift cursor...

Страница 9: ...se retaining arm 3 5 SWING retaining arm 3 fully open 6 REMOVE sealing blanks on connector manifold Installation of Protek cartridge Protek L1 or L2 acceptance screen 7 REMOVE Protek pre treatment cartridge LC175 LC177 from packaging 8 REMOVE sealing plugs from inlet and outlet ports 9 WET O rings located inside the ports 10 ALIGN Protek cartridge 4 on guide rails and SLIDE into place until it is ...

Страница 10: ...ther water purity an additional stage of purification can be added in the form of a purification cartridge Labpure L1 and a Micro Filter See Operator Manual for installation instructions or refer to Technical Support 1 Wait for to be displayed 2 PRESS 4 3 System purge 1 PRESS the system now enters an automatic purge phase After a maximum of 60 minutes the system will exit the purge phase regardles...

Страница 11: ...S FOLLOW RECOMMENDATIONS FOUND IN THE MANUFACTURERS MATERIAL SAFETY DATA SHEET AND ANY APPLICABLE OSHA STANDARDS FOR THE CHEMICAL BEING USED WARNING LABEL THE MACHINE WITH APPROPRIATE WARNING SIGNS SUCH AS DO NOT USE CONTAINS STERILANT NOT PROVIDED Minncare Cold Sterilant is a Peracetic and Hydrogen Peroxide based solution Refer to the Minncare label for additional information and follow all appli...

Страница 12: ...tization 4 PRESS to enter RO sanitization Large white lid removal 5 UNSCREW lid on the left of the unit 6 MEASURE out 50ml of Minncare Cold Sterilant and SLOWLY POUR into the break tank 7 REPLACE lid Confirmation the chemical is added 8 PRESENT GREEN sanitization key to confirm that the chemical has been added ...

Страница 13: ... 9 The system now enters a period of automatic sanitization lasting approximately 1 hour Once the step within the sanitization is complete the unit will stop and show date re set Set sanitization reminder 10 PRESS to accept the sanitization reminder the unit will go into process and continue to fill the recirculation reservoir ...

Страница 14: ...Sanitization 1 ISOLATE application and ensure the process is OFF 2 PRESENT GREEN sanitization key to reader Sanitization selection screen 3 PRESS to enter sanitization process Enter RO sanitization 4 PRESS to show recirculation sanitization icon 5 PRESS sanitization to select recirculation Present the sanitization key ...

Страница 15: ...ESS to allow the system to empty fill the reservoir The system will stop when the level reaches approximately 10 liters Small white lid removal 8 UNSCREW lid on the right of the unit 9 MEASURE out 250ml of Minncare Cold Sterilant and SLOWLY POUR into the main reservoir 10 REPLACE lid Confirmation the chemical is added 11 PRESENT GREEN sanitization key to confirm that the chemical has been added ...

Страница 16: ... the unit will go straight into process and continue to fill the recirculation reservoir Set sanitization reminder 13 REFIT Labpure L1 LC187 14 PRESS to accept the sanitization reminder The system is now commissioned and ready for operation Note The automated rinse cycles are designed to remove the chemical sanitant from the system but for critical applications a water test is advised prior to use...

Страница 17: ...MENDATIONS FOUND IN THE MANUFACTURERS MATERIAL SAFETY DATA SHEET AND ANY APPLICABLE OSHA STANDARDS FOR THE CHEMICAL BEING USED WARNING LABEL THE MACHINE WITH APPROPRIATE WARNING SIGNS SUCH AS DO NOT USE CONTAINS STERILANT NOT PROVIDED Refer to the EfferSan label for additional information and follow all applicable directions for use on the manufacturer s label in conjunction with the following ins...

Страница 18: ... Tablet cutter for cutting EfferSan tablets 4 PRESS to enter RO sanitization CAUTION ENSURE THE CORRECT DOSAGE OF CHEMICAL IS USED Large white lid removal 5 UNSCREW lid on the left of the unit 6 CUT one EfferSan tablet into four quarters using tablet cutter 7 INSERT of EfferSan tablet into the break tank 8 REPLACE lid ...

Страница 19: ...e chemical has been added RO sanitization process steps 10 The system now enters a period of automatic sanitization lasting approximately 1 hour Once the step within the sanitization is complete the unit will stop and show date re set Set sanitization reminder 11 PRESS to accept the sanitization reminder the unit will go into process and continue to fill the recirculation reservoir ...

Страница 20: ...Sanitization 1 ISOLATE application and ensure the process is OFF 2 PRESENT GREEN sanitization key to reader Sanitization selection screen 3 PRESS to enter sanitization process Enter RO sanitization 4 PRESS to show recirculation sanitization icon 5 PRESS sanitization to select recirculation Present the sanitization key ...

Страница 21: ...rvoir The system will stop when the level reaches approximately 10 liters CAUTION ENSURE THE CORRECT DOSAGE OF CHEMICAL IS USED Small white lid removal 8 UNSCREW small lid on the right of the unit 9 CUT one EfferSan tablet into two halves using tablet cutter 10 INSERT Half of an EffenSan tablet into the break tank 11 REPLACE lid Confirmation the chemical is added 12 PRESENT GREEN sanitization key ...

Страница 22: ... the unit will go straight into process and continue to fill the recirculation reservoir Set sanitization reminder 14 REFIT Labpure L1 LC187 15 PRESS to accept the sanitization reminder The system is now commissioned and ready for operation Note The automated rinse cycles are designed to remove the chemical sanitant from the system but for critical applications a water test is advised prior to use...

Страница 23: ...wing modes of operation Night service Operational day Continuous Recirculation 24 7 Sanitization Procedure CENTRA R 60 120 US is supplied set up to operate on Monday to Friday between the hours of 07 30 to 18 00 Refer to Operator Manual supplied on CD for instructions on how to adjust the times of operation or to switch to continuous recirculation ...

Страница 24: ...S ELGA LabWater Lane End Industrial Park High Wycombe Bucks HP14 3BY UK Tel 44 0 203 567 7300 Fax 44 0 203 567 7205 E mail techsupport elgalabwater com For the address of your nearest ELGA LabWater Sales and Service office visit the country list on our website http www elgalabwater com or contact ELGA at the number above ...

Отзывы: