2. IMPORTANT HEALTH AND SAFETY INSTRUCTIONS
WARNING!
WARNINGS ARE GIVEN WHERE FAILING TO OBSERVE
THE INSTRUCTIONS COULD RESULT IN INJURY OR
FATALITY.
CAUTION!
Cautions are given where failure to observe the
instructions could result in damage to the equipment,
associated equipment and processes.
2.1 Environment
CAUTION!
Clean dry indoor. Temp 5-40°C.
Humidity max 80% non-condensing.
Failure to follow the environmental specification could
result in damage to the system.
WARNING!
UNIT TO BE PLACED IN A ROOM WHICH IS A MINIMUM
OF 10m³ AND VENTILATED TO PROVIDE AIR MOVEMENT
OF SIX AIR CHANGES EVERY HOUR.
WARNING!
IF CRITICAL ALARMS ARE ACTIVATED. ACTIVATE THE
BYPASS VALVES, AND ISOLATE THE UNIT FROM THE
MAINS ELECTRICAL SUPPLY, AND CONTACT YOUR
SERVICE PROVIDER.
2.2 Electricity
The appliance coupler (mains lead) or power supply connected to the rear of the unit can
be removed to isolate the power supply. If access to this is restricted then it is
recommended that access to the supply socket is easily available to disconnect the
electrical supply.
WARNING!
ONLY USE THE APPLIANCE COUPLER (MAINS LEAD) AND
POWER SUPPLY PROVIDED.
THE USE OF THESE WILL ENSURE ADEQUATE EARTH
PROTECTION IS PROVIDED.
IF THE EQUIPMENT IS USED IN A MANNER NOT
SPECIFIED BY ELGA VEOLIA, THE PROTECTION
PROVIDED BY THE EQUIPMENT MAY BE IMPAIRED,
POSITION THE POWER SUPPLY SO THAT IT CANNOT
COME INTO CONTACT WITH WATER.
MEDICA biox MANU40669 Version 03 Page 7