User Manual RPS/T 20kVA 62000587_02_
10
motor with starting capacitor, or in all those cases where a different phase displacement than
three phase is required.
The resolution of phase setting is 0,088° (12 bits on 360°).
1.1.3.
User interface
RPS/T can be managed from an host computer control, thus allowing to perform tests
automatically.
RPS/T allows various usage selections: working range, wires drop compensation, working
frequency, synchronization of the output frequency with the power line.
RPS/T gives the user clear information on the status of the output.
It is monitored both the set voltage and the set frequency and the output voltage is read with a
precision of 0.3% f.s.
The user is also warned in case of overcurrent obtainable by the RPS/T, or in case of high loss in
the wires, that should not exceed 5% of the set voltage.
We underline again that RPS/T automatically limits the maximum allowed current, avoiding
damages to the equipment; in that in this case it is not guaranteed the precision of the output
waveform neither the accuracy of the output voltage.