2. Specifications

2-1. General Specifications

Display

3 3/4 digit LCD with a max. reading of 4000.

Range Control

Manual range control.

Polarity

Automatic negative polarity indication.

Zero Adjustment

Automatic.

Overrange Indication

“OL” on LCD is displayed.

Low Battery Indication

“        ” sign on LCD readout.

Auto Power Off

Approximately 15 minutes.

Safety Standards

EMC/LVD.

The meter is up to the

standards of IEC1010 Pollution Degree 2,
Overvoltage Catagory II.

Operation Temperature

32

O

F to 104

O

F (0

O

C to 40

O

C)

less than 85% relative humidity

Storage Temperature

–4

O

F to 140

O

F (–20

O

C to 60

O

C)

less than 95% relative humidity

Power

9V battery.

Dimensions

7 5/8” (H) x 3 9/16” (W) x 1 3/4” (D)
(w/o holster)

Weight

Approximately 0.66 lb. (w/o holster)

Accessories

Safety Test Leads - 1 pair
Operator’s Manual - 1 pc.
Holster - 1 pc.

2-2. Electrical Specifications

Accuracy is + (% or r number in last digit)     at 23 + 5

O

C, <75% RH.

DC Voltage

400mV, 4V, 40V, 400V

+(0.5% + 3)

1000V

+(0.8% + 3)

Impedance

10M

W

Max. input: 250V~ on 400mV range; 1000V~ on all other ranges.

AC Voltage

4V, 40V, 400V

+(0.8% + 3)

400mV, 750V

+(1.2% + 3)

Impedance

10M

W

Max. input: 250V~ on 400mV range; 1000V~ on all other ranges.

Resistance

400

W

, 4k

W

, 40k

W

, 400k

W

, 4M

W

+(1% + 3)

40M

W

+(2% + 3)

400M

W

+(5% + 10)

Overload Protection

250V DC/ACrms

DC Current

40

m

A, 400

m

A, 4000

m

A

+(1.2% + 3)

40mA, 400mA

+(1.5% + 3)

10A

+(2% + 5)       

Caution: Max. operation time 15s.

Overload Protection

Fast 0.5A/250V, 10A/250V fuse.

AC Current

40

m

A, 400

m

A, 4000

m

A

+(1.5% + 3)

40mA, 400mA

+(2% + 3)

10A

+(2.5% + 5)       

Caution: Max. operation time 15s.

Overload Protection

Fast 0.5A/250V, 10A/250V fuse.

Frequency Response

40 - 400Hz

Capacitance

4nF

+(5% + 10)

40nF

+(3% + 10)

400nF, 4

m

F, 40

m

F

+(2% + 5)

200

m

F

+(4% + 5)

Overload Protection

250V DC/ACrms

Frequency

Frequency 10Hz - 10MHz

+(0.1% + 5)

Sensitivity

Sine wave 0.6Vrms

Overload Protection

250V DC/ACrms

Duty Cycle

Duty Cycle 0.1% - 99.9%

+(2% + 2) Frequency lower than 10kHz

Sensitivity

Sine wave 0.6Vrms

Overload Protection

250V DC/ACrms

Diode Test

Test Current: 1 +0.6mA

Test Voltage: Approx. 1.5V

Overload Protection

250V DC/ACrms

Continuity Test

Audible Indication

Less than 120

W

approx.

Overload Protection

250V DC/ACrms

h

FE

Test

Ib

10mA

Vce

2.5V approx.

Test Range

0 - 1000

Logic Test

TTL        Logic 1: 2.5V +0.8V

Logic 0: 0.8V +0.5V

CMOS    Logic 1: 4.0V +1.0V

Logic 0: 2.0V +0.5V

Frequency Response: 20MHz
Detectable Pulse Width: 25ns min

-4-

-5-

+

Max. voltage from “V

W

mA”

socket to “COM” socket is
1000V~ from “COM” socket
to earth is 500V~.

!

!