E-4
ElektroPhysik
Accessories, Safety notes
2.
Safety notes
-
PoroTest® 7 is designed for safe use. When ob-
serving the safety notes, operation of the
instrument is absolutely risk free to the operator.
Current and capacitance limits conform to the
German standards DIN EN 61010/ Part 1 (VDE
0411/Part 1).
-
Persons with cardiac pacemakers should consult
their physician about the possibility of interference
with the instrument.
-
Caution!
Never use unearthed equipment!
B e f o r e
use, make sure both, PoroTest® 7 and the test
object are well earthed (see section 3.)
-
Only use rechargeable NiMH C-cells. Non
rechargeable batteries pose an explosion hazard
if on charge!
-
After the control unit has been switched off,
discharge the high-voltage generator to earth.
-
Do not carry out high voltage discharge to the
instrument or to the connecting cable between
instrument and probe.
1.4
Accessories
·
Aluminium case
·
Brush, rolling spring, ring and right-angle
electrodes in various designs, silicon-rubber
electrode, 200 mm (7.8 ft.) width
·
3-pin plug, cable-free, directly plugs into the
integrated signal contact (make/break
contact).The signal contact is automatically
actuated if a pore is detected and may be used
for an external signal devices.
·
Earthing rod
·
Earthing/equipotential cable, 10 m (32 ft.) length
·
Connecting cable for control unit and high voltage
probe in special lengths: 5m (16 ft.), 10m ( 32 ft.)
·
Non-destructive coating thickness gauges, help
you to adjust test voltage specific to coating
thickness