Measurement
ElektroPhysik
E-6
【
6.2
】
Measurement
There are two ways to access to measurement status, which are:
1.
Insert the standard probe and switch on the instrument to go to the
measuring status.
2.
During menu operation, press “
ESC
” to return back to the measuring
status
.
【
6.3
】
Calibration
It will cause error during the primary stage of usage and operating, this
caused by the following three aspects:
1.
The probe itself or the temperature variation
2.
System error caused by the match of the instrument and the probe.
3.
Calculation error caused by the sound velocity set in the instrument is
different from that of the actual material. In order to eliminate the possible
error, the instrument provides three solutions to resolve it.
Press
key, and then press
or
to select the options.
〖
6.3.1
〗
Auto calibration
Using different probe, wear out probe and the environment temperature
will all cause measurement error, for these factors, the “Auto calibration”
function is available.