PS 2000 B Triple Series

3.4.7  Control panel (HMI) lock

In order to avoid the accidental alteration of a value during manual operation, the rotary knobs and the buttons can 

be locked so that no action is accepted without prior unlocking.

 

How to lock the HMI

1. 

While the DC output (1 or 2) is switched off, press button 

Settings

.

2. 

In the menu use the left knob to navigate to 

HMI Setup

 and press the knob.

3. 

In menu 

HMI Setup

 use the knob to select and configure the lower four parameters related to the HMI lock. 

For details about the single parameter see section 

«3.4.3.3. Menu “HMI Setup”».

4. 

The HMI lock is activated with 

Lock HMI? = Yes

 and leaving the menu. The active lock is indicated by 

symbol 

.

Alternatively to the simple lock, which can be unlocked very easily by every person and thus offers no protection 

against intentional misuse, a PIN can set up and activated, which then is requested to be entered every time the 

HMI is going to be unlocked.

 

How to lock the HMI with PIN

Don’t activate the PIN lock if you are unsure about the current PIN! It can be changed, but only 

if the current PIN is entered. Resetting the device via remote command will also reset the PIN 

to the default 0000.

1. 

In the menu where you configured the HMI lock select parameter 

Enable PIN

 and set the parameter to 

Yes

 

with the right-hand knob.

2. 

In order to change the PIN prior to activation select 

Change PIN

 and the left knob to access the next screen 

where you are requested to enter the former PIN 1x and the new PIN 2x and confirm every step with the 

left knob.

3. 

The HMI lock is activated with 

Lock HMI? = Yes

 and leaving the menu. The active lock is indicated by 

symbol 

.

If an attempt is made to alter something whilst the HMI is locked, a requester appears in the display asking if the 

lock should be disabled.

 

How to unlock the HMI

1. 

Turn one of the rotary knobs or press any button (except for 

On/Off

 when lock mode 

On/Off possible

 has 

been set).

2. 

This request pop-up will appear: 

.

3. 

Unlock the HMI by pressing the left knob within 5 seconds, otherwise the pop-up will disappear and the HMI 

remains locked. In case the additional PIN code lock

 

has been activated in the menu “

HMI Setup

”, another 

requester will pop up, asking you to enter the PIN before it finally unlocks the HMI.

Содержание 39 200 120

Страница 1: ...Elektro Automatik PS 2000 B TFT Triple DC Power Supply Operating Manual Doc ID PS2TEN Revision 14 Date 11 2020...

Страница 2: ......

Страница 3: ...res before installation and use 14 2 3 2 Preparation 14 2 3 3 Installing the device 14 2 3 4 Connection to DC loads 15 2 3 5 Grounding of the DC outputs 15 2 3 6 Connecting the USB port 15 3 OPERATION...

Страница 4: ...as well as general notices in this document are shown in a box with a symbol as follows Symbol for a life threatening danger Symbol for general safety notices instructions and damage protection bans S...

Страница 5: ...B Construction Version B Second generation Maximum current of the device in Ampere Maximum voltage of the device in Volt Special marking 3 Three outputs triple Series 2 Series 2000 Type identification...

Страница 6: ...ulations must be applied when plugging interface cards or modules into the relative slot Interface cards or modules may only be attached or removed after the device is switched off It s not necessary...

Страница 7: ...he responsibilities for operation maintenance and cleaning of the equipment ensure that all employees who use the equipment have read and understood the manual Furthermore the users are to be regularl...

Страница 8: ...pair of DC output sockets and is called output 2 The upper pair of DC output sockets form output 3 The meaning of the signals on the display of output 1 and 2 is as follows Signal OT OverTemperature...

Страница 9: ...P 24 mVRMS Output 1 2 Current Adjustment range 0 6 A 0 10 A 0 3 A 0 5 A Accuracy 1 at 23 5 C 0 3 IMax 0 3 IMax 0 3 IMax 0 3 IMax Load regulation at 0 100 UOUT 0 15 IMax 0 15 IMax 0 15 IMax 0 15 IMax O...

Страница 10: ...Automatik GmbH Helmholtzstr 31 37 41747 Viersen Germany Fon 49 2162 3785 0 Fax 49 2162 16230 www elektroautomatik de ea1974 elektroautomatik de PS 2000 B Triple Series 1 8 4 Views Figure 1 Front side...

Страница 11: ...alue or adjustment of parameter value in the menu Push selection of the decimal position cursor of the currently assigned value 7 Pushbutton On Off Used to switch the DC output oon or off unless the H...

Страница 12: ...es For these devices the following accessories are available Multi Control license Ordering numbers Single license 33 100 229 5 pack of licenses 33 100 230 The device is delivered with an USB stick th...

Страница 13: ...tatus display lower part This area displays various status texts and symbols Display Description The HMI is locked The HMI is unlocked Remote The device is under remote control via USB Alarm xxx Alarm...

Страница 14: ...3 1 Safety procedures before installation and use Before connecting to the mains ensure that the connection is as shown on the product label Overvoltage on the AC supply can cause equipment damage In...

Страница 15: ...ding of the DC outputs The metal socket in proximity to output 1 is internally wired to PE and can be used to ground one of the DC output poles of all three outputs by direct connection or it can be u...

Страница 16: ...e Voltage regulation is also called constant voltage operation CV The DC output voltage of a power supply is held constant on the adjusted value unless the output current reaches the adjusted current...

Страница 17: ...will be in constant voltage mode If however the actual current reaches the set value of current the device will switch automatically to current limiting While the DC output is switched on and constan...

Страница 18: ...ltage alarm OVP will switch off the DC output and can occur if the power supply itself as a voltage source generates an output voltage higher than set for the overvoltage alarm threshold OVP 0 110 UNo...

Страница 19: ...or a blackout doesn t leave the device enough time to store the latest values It s practically powered down immediately Thus the device stores the set values and all settings automatically when they...

Страница 20: ...left HMI Element Description In right HMI Language Selection of the display language between German and English Default setting English No Backlight The choice here is whether the backlight remains Al...

Страница 21: ...bs 1 With the main display being shown it means no menu mode is active turn the left hand knob to adjust the output voltage and the right hand knob to adjust the output current This doesn t depend on...

Страница 22: ...How to lock the HMI with PIN Don t activate the PIN lock if you are unsure about the current PIN It can be changed but only if the current PIN is entered Resetting the device via remote command will a...

Страница 23: ...ivated tracking is saved along with other device settings and thus instantly restored after powering the device Even in tracking mode the two outputs can have separate alarms which would always lead t...

Страница 24: ...t used the serial current is limited by the output with the lower current set value Series connection of multiple devices isn t permitted because the total voltage could exceed the voltage withstand o...

Страница 25: ...considered completely differently to the other outputs It s meant to be an auxiliary output with a fixed voltage which is barely adjustable ca 3 to ca 6 V with a trimmer That trimmer is placed between...

Страница 26: ...can try to acknowledge and delete the alarm by pressing button On Off In order to acknowledge an alarm during remote control see the programming guide These device alarms are configurable Alarm Meani...

Страница 27: ...detail as possible is attached if shipping destination is abroad the necessary customs papers are attached 4 2 1 Replacing a defect mains fuse The device is protected by a fusible which is inside a f...

Страница 28: ...EA Elektro Automatik GmbH Co KG Development Production Sales Helmholtzstra e 31 37 41747 Viersen Germany Fon 49 2162 37 85 0 Telefax 49 2162 16 230 ea1974 elektroautomatik de www elektroautomatik de...

Отзывы: