Basic measurements
8
NM21283A-A
49
8.7 FFT measurement
The job “FFT” is for measuring power spectra of the channels. This can
be used to examine the external disturbances and noise characteristics in
more detail compared to the “Noise” measurement. The spectrum estima-
tion uses a Hanning window and non-overlapping averaging.
The parameters affecting the FFT measurements are shown in table below.
The Time buffer size defines the length of the time trace that is actually
measured. From the beginning of each time trace, FFTsize samples are
taken and transformed into Spectral noise density per unit band (1/Hz).
The length of the time buffer should be at least the length of the Fourier
transform. In order to get better statistics, several spectra can be averaged.
The number of averages is controlled by parameter FFTaverages. It is pos-
sible to apply SSP noise reduction to the data before calculating the FFT
in order to diminish the contribution of external disturbances.
The last four parameters in Table 8.3 define two bands, one for numerical
estimation of the low frequency noise and another for estimating the white
noise level. The noise power in these bands can be shown as a bar dia-
gram.
Parameter
Type
Description
TimeBufSize
integer
Time buffer length. [samples]
FFTsize
integer
FFT transformation length. [samples]
FFThpf
float
High pass filter setting. [Hz]
FFTlpf
float
Low pass filter setting. [Hz]
FFTsfreq
float
Sampling frequency. [Hz]
FFTaverages
integer
Number of averages.
FFTssp
0 or 1
Is SSP used?
FFTlowStart
float
Start of low freq. estimation range. [Hz]
FFTlowEnd
float
End of low freq. estimation range. [Hz]
FFTwhiteStart
float
Start of white noise estimation range.[Hz]
FFTwhiteEnd
float
End of white noise estimation range. [Hz]
Table 8.3 Parameters affecting FFT measurements
Содержание NM21283A
Страница 1: ...Sensor Tuner User s Guide Software Version 3 0 for Elekta Neuromag March 2005...
Страница 6: ...iv NM21283A A...
Страница 8: ...2 NM21283A A...
Страница 10: ...4 NM21283A A...
Страница 14: ...1 User interface basics 8 NM21283A A...
Страница 30: ...5 Description of the menus 24 NM21283A A...
Страница 32: ...26 NM21283A A...
Страница 46: ...7 Manual tuning 40 NM21283A A...
Страница 48: ...42 NM21283A A...
Страница 66: ...11 Miscellaneous 60 NM21283A A...