6.2.6. The item
“Calibr by air”
sets the
level
“0”
for all channels except oxygen
according to the current state of the
environment. It establishes a normal
concentration of oxygen in the
environment - 20.95 ppm. Use only in the
“clean”
zone, knowing in advance that
there are no additional impurities of
suspended gases in the environment, etc
.
Choose by butoon
«Yes» and press
Air calibration done.
6.2.7. When selecting
“Calibration”
, a
window for selecting a channel for
calibration will open
After selecting a channel,
the “action
selection”
window appears
Where the channel number and type of
gas are indicated in the status window.