ELECTROMATIC TI-25M-MMX Скачать руководство пользователя страница 5

– 5 –

3.4 The  Keypad

The TI-25M-MMX interacts with the operator through the
membrane keypad and the LCD display. The functions of
the various keys on the keypad are detailed below, followed
by an explanation of the display and its various symbols.

This key is used to turn the TI-25M-MMX on and off. When the 
gauge is turned ON, it will first perform a brief display test by 
illuminating all of the segments in the display. After one second, 
the gauge will display the internal software version number. After 
displaying the version number, the display will show “0.000” (or 
“0.00” if using metric units), indicating the gauge is ready for use. 
The TI-25M-MMX is turned OFF by pressing the ON/OFF key. 
The gauge has a special memory that retains all of its settings even 
when the power is off. The gauge also features an auto-power 
down mode designed to conserve battery life. If the gauge is idle 
for 5 minutes, it will turn itself off.

The PRB-0 key is used to “zero” the TI-25M-MMX in much the 
same way that a mechanical micrometer is zeroed. If the gauge is 
not zeroed correctly, all of the measurements that the gauge makes 
may be in error by some fixed value. Refer to page 10 for an 
explanation of this important procedure.

The CAL key is used to enter and exit the TI-25M-MMX’s

calibration mode. This mode is used to adjust the sound-velocity 

value that the TI-25M-MMX will use when calculating thickness. 
The gauge will  either calculate the sound-velocity from a sample 
of the material being measured, or allow a known velocity value to 
be entered directly. Refer to page 11 for an explanation of the two 
CAL functions available.

The IN/MM key is used to switch back and forth between English 
and metric units. This key may be used at any time, whether the 
gauge is displaying a thickness (IN or MM) or a velocity value 
(IN/ms or M/s).

The UP arrow key has two functions. When the TI-25M-MMX is 
in calibration mode, this key is used to increase numeric values on 
the display. An auto-repeat function is built in, so that when the 
key is held down, numeric values will increment at an increasing 
rate. When the TI-25M-MMX is not in calibration mode, the UP 
arrow key switches the SCAN measurement mode on and off. 
Refer to page 15 for an explanation of the SCAN measurement 
mode.

ALRM

ON

OFF

IN

M M

DUAL

M ULTI

PRB

SEND

CAL

SCAN

0

ON

OFF

PROBE

ZERO

CAL

INCH

MM

SCAN

– 24 –

1 1 .0  S

PECI FI CAT I ON S

Range 
TI-25M-MMX* 

Pulse-Echo Mode (Pit & Flaw Detection):
0.040–6.000" (1.01–152 mm)
Echo-Echo Mode (Through Paint & Coatings):
0.100–1.0" (2.54–25.4 mm). With up to 0.040" of coating.

TI-25M-MMX-EXT* Pulse-Echo Mode (Pit & Flaw Detection):

0.100–10.000" (2.54–254 mm)
Echo-Echo Mode (Through Paint & Coatings):
0.200–5.0" (5–127 mm). With up to 0.080" of coating.

Resolution

.001" (0.01 mm)

Display 

4

1

2

-Digit, 0.5" Backlit LCD

Velocity Range

0.0492 to .3937 in./µs. (1250 to 10,000 meters/second)

Probe 
TI-25M-MMX

5 MHz, 0.25" Dia. (6.35 mm), High Damp

TI-25M-MMX-EXT

3.5 MHz, 0.5" Dia. (12.70 mm), High Damp

Probe Wearface

PEEK (Polyethylethylkeytone)

Cable 

4 ft. (1.2 m) waterproof cable with non-polarized, 
quick-disconnect connectors. Optional lengths up to
100 ft. (30 m).

Probe Zero 

Steel plate built into battery cover, appriximate thickness 

Test Plate

of 0.416" (10.57mm)

Temp. Limits 

Ambient: –20 to 120 °F (–30 to 50 °C) 
Material:

0 to 200 °F (–20 to 100 °C)

Special high temperature probes are optionally available.

Battery Type 

Two AA batteries

Battery Life 

200 hours

Weight 

7 ounces (196 g)

Size

2.4 x 4.5 x 1.25" (65 x 114 x 35 mm)

Accessories 
Included

Probe/cable assembly. 4oz. bottle of coupling fluid, 
NIST Calibration Certificate, 2 AA batteries, operating 
instructions,hard-plastic carrying case.

Warranty 

Gauge: 5 years 
Probe: 90 days

*Measuring Range indicated is for steel. Actual range for other materials will
vary based upon the material’s sonic velocity and attenuation.

Содержание TI-25M-MMX

Страница 1: ...LECTROMATIC All returns for warranty or non warranty repairs and or replacement must be authorized by ELECTROMATIC in advance with all repacking and shipping expenses to the address below to be borne by the purchaser THE FOREGOING WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES EXPRESSED OR IMPLIED INCLUDING BUT NOT LIMITED TO THE WARRANTY OF MERCHANTABILITY AND FITNESS FOR ANY PARTICULAR PURPOSE OR A...

Страница 2: ...iece of known thickness Ideally this sample material should be a part of the same piece being measured or at least from the same lam ination batch By calibrating to each test piece individually the effects of varia tion of sound velocity will be minimized An additional important consideration when measuring laminates is that any included air gaps or pockets will cause an early reflection of the ul...

Страница 3: ...ecreases In most applications with surface temperatures less than about 200 F 100 C no special procedures must be observed At temperatures above this point the change in sound velocity of the material being measured starts to have a noticeable effect upon ultrasonic measurement At such elevated temperatures it is recommended that the user perform a calibration procedure refer to page 11 on a sampl...

Страница 4: ...reezing Point 20 F Vapor Pressure N A Evaporation Rate N A Specific Gravity 1 0Z Solubility in Water complete pH 7 35 7 9 Acoustic Imp 1 726x106 Vapor Density N A Appearance and Odor water white opaque gel bland odor Section 4 Fire and Explosive Hazard Data Flash Point none Upper Exposure Limit none Lower Exposure Limit none Special Fire Fighting Procedures N A Extinguishing media N A Unusual Fire...

Страница 5: ...M or a velocity value IN ms or M s The UP arrow key has two functions When the TI 25M MMX is in calibration mode this key is used to increase numeric values on the display An auto repeat function is built in so that when the key is held down numeric values will increment at an increasing rate When the TI 25M MMX is not in calibration mode the UP arrow key switches the SCAN measurement mode on and ...

Страница 6: ...lay numeric values as well as occasional simple words to indicate the status of various settings When the TI 25M MMX is displaying thickness measure ments the display will hold the last value measured until a new measure ment is made Additionally when the battery voltage is low the entire dis play will begin to flash When this occurs the batteries should be replaced These eight vertical bars form ...

Страница 7: ...ng the probe supplied with the TI 25M MMX the user should perform a Probe Zero as described in Section 4 2 The Probe Zero Test Plate is used for this task It is located on the top edge of the gauge as shown in the photo below It also serves as the battery compartment cover Note The thickness of this plate is not important and it should not be used as a Calibration Test Plate A precision 4 step Tes...

Страница 8: ...the factory calibrated for steel with an acoustic velocity of 0 2330 IN µs 5920 M s To measure the thickness of any other material the calibration will have to be changed by adjusting the acoustic velocity to the appropriate value for the specific material being measured To determine the proper acoustic velocity for the non steel material refer to the Acoustic Velocity Table section 10 1 After det...

Страница 9: ... this happens the TI 25M MMX may perform a measurement through this couplant film resulting in a measurement that is larger or smaller than it should be This phenomenon is obvious when one thickness value is observed while the transducer is in place and another value is observed after the transducer is removed In addition measurements through very thick paint or coatings may result in the paint or...

Страница 10: ...lse echo mode by pressing the Dual Multi key to toggle the modes Note The Probe Zero feature is not used in Echo Echo Thru Paint mode and has been disabled If the PRB 0 key is pressed while in this mode nO followed by Prb0 will be displayed 19 10 0 TRANSDUCER SELECTION The TI 25M MMX is inherently capable of performing measurements on a wide range of materials from various metals to glass and plas...

Страница 11: ...tting the sound velocity described in the following pages Note Although the TI 25M MMX has a thru paint coating feature one and two point calibrations must be performed on material with the paint or coating removed Failure to remove the paint or coating prior to calibration will result in a multi material velocity calculation that may be different from the actual material velocity intended to be m...

Страница 12: ...M symbol is flashing press the CAL key again so that the IN ms or M s symbols are flashing 3 Use the UP and DOWN arrow keys to adjust the displayed velocity up or down until it matches the sound velocity of the material to be measured 17 8 0 DUAL MULTI MODE Often times users and inspectors in the field are faced with coated materials such as pipes and tanks Typically inspectors will need to remove...

Страница 13: ...upled repeat step 4 6 Press the CAL key The IN or MM symbol should begin flashing 7 Use the UP and DOWN arrow keys to adjust the displayed thickness up or down until it matches the thickness of the sample piece 8 Press the Probe key The display will flash 1OF2 Repeat steps 3 through 8 on the second calibration point The TI 25M MMX will now display the sound velocity value it has calculated based o...

Страница 14: ...emoved from the sample after a measurement the last reading will be retained on the display b Occasionally a small film of couplant will be drawn out between the probe and the surface as the probe is removed When this happens the TI 25M MMX may perform a measurement that is larger or smaller than it should be This phenomenon is obvious when one thickness value is observed while the probe is in con...

Страница 15: ...ng Instructions CHECK LINE BY ELECTROMATIC Includes special version Model TI 25M MMX EXT MMX310 Distributed by ABQ Industrial LP USA Tel 1 281 516 9292 888 275 5772 eFax 1 866 234 0451 Web https www abqindustrial net E mail info abqindustrial net ...

Отзывы: