– 8 –

3.0

TI-007DL MODES AND FEATURES

The Back Light

The back light of the TI-007DL can be set to three different positions on/off/auto.  
The following steps outline the procedure for changing the settings of the back light:

1. Press ON/OFF key to power up the TI-007DL.

2. Press the MODE key to activate features and settings.

3. Press the UP/DOWN arrow keys to scroll to the LItE symbol.

4. Press the SEND key to toggle the status of the back light  on/off/auto.

5. Press the MODE key once again to return to measurement mode.

Scan Mode

While the TI-007DL excels at making single point measurements, it is sometimes
desirable to examine a larger region, searching for the thinnest point. The TI-007DL
includes a feature, called Scan Mode, which allows it to do just that.

In normal operation, the TI-007DL performs and displays four measurements every
second, which is quite adequate for single measurements. In Scan Mode, however, the
tool performs eight measurements every second, but does not display them. While the
transducer is in contact with the material being measured, the TI-007DL is keeping
track of the lowest measurement it finds. The transducer may be “scrubbed” across a
surface, and any brief interruptions in the signal will be ignored. When the transducer
loses contact with the surface for more than a second, the TI-007DL will display the
smallest measurement it found.

When the TI-007DL is in measurement mode, press the MODE key to activate the fea-
tures and settings. The display will begin flashing ALAr (alarm), followed by the sta-
tus - on/off. Use the UP and DOWN arrow keys to scroll to SCAn mode.  Press the
SEND key to toggle scan mode on/off. While scanning, the display will show a mov-
ing series of dashes instead of a thickness value. When the transducer is removed from
the material being scanned, the TI-007DL will (after a brief pause) display the smallest
measurement it found.

Alarm Mode

The Alarm Mode feature of the TI-007DL allows the user to set an audible and visual 
Hi /Lo parameter when taking measurements. If the measurement falls below or above
the Hi /Lo limits, set by the user, a red light will be illuminated on the front panel of
the gauge and an audible indication (beep) will sound. This improves the speed and
efficiency of the inspection process by eliminating constant viewing of the actual read-
ing displayed. The following procedure outlines how to enable and set up this feature:

Using the Beeper

1. Press ON/OFF key to power up the unit

2. Press MODE key to activate features and settings.

3. Press the UP and DOWN arrow keys to scroll to bEEP.

6. Press the UP/DOWN arrow keys to scroll to the flashing CLr/F-05 (the file 

selected by the user) symbol.

7. Press the SEND key to select the clear file option.  The symbol (CLr?) will be 

displayed.

8. Press the CLR key to confirm and clear the contents of the entire file.

9. Press the MEM key, at any time, to exit the data logging functions and return 

to measurement mode.

Clearing all Files

1. Press the ON/OFF key to power up the TI-007DL.

2.  Immediately press the CLR key.  CLr? will be displayed.

3. Press the CLR key once again to clear all files.

Sending All Files to a Computer

At the end of the inspection process, or end of the day, the user may require the 
readings be transferred to a computer.  The following steps outline this procedure:

1. Refer to the section on Connecting to a Computer, page 11, before proceeding.

2. Press the ON/OFF key to power up the TI-007DL.

3. Press the MEM key to activate the data logging functions and settings.

4. Press the UP/DOWN arrow keys to scroll to the SEND/ALL symbol flashing 

on the display.

5. Press the SEND key to send all data files to the computer.

6. Press the MEM key to exit the data logging functions and return to 

measurement mode.

Printing a File

The user may wish to print an individual file to a serial printer or computer.  
A file can, very simply, be printed to a communications program on a PC 
(i.e., Windows 3.1 - Terminal, Window 95/98 - HyperTerminal), and then printed.  
The procedure for printing a file is outlined below:

1.  Refer to the section on Connecting to a Computer, page 11, before proceeding.

2. Press the ON/OFF key to power up the TI-007DL.

3. Press the MEM key to activate the data logging functions and settings.

4. Press the SEND key to enter file setup.

5. Press the UP/DOWN arrow keys to scroll to the file to be printed (F-01, F-05, etc.).

6. Press SEND, once again, to select the file to be printed. The display will flash 

the FILE/F-05 ( The file selected ) symbol.

– 13 –

Содержание TI-007DL

Страница 1: ... Modes 9 4 1 Differential Mode 4 2 S 232 Port 5 0 RS 232 Serial Port Data Logger 11 5 1 Connecting To A Computer 5 2 Using The Data Logger 5 3 Clearing A Storage Location 5 4 Clearing An Entire File 5 5 Clearing All Files 5 6 Sending Files To A Computer 5 7 Printing Files 6 0 Transducer Selection 15 7 0 Appendix A Product Specification 16 8 0 Appendix B Application Notes 17 9 0 Appendix C Sound Ve...

Страница 2: ...h manifests itself within one year from the sale of such product shall be remedied by repair or replacement of such product at ELECTROMATIC s option except where unauthorized repair disassembly tampering abuse or misappli cations has taken place as determined by ELECTROMATIC All returns for warranty or non warranty repairs and or replacement must be authorized by ELECTROMATIC in advance with all r...

Страница 3: ... on the display An auto repeat function is built in so that when the key is held down numeric values will increment at an increasing rate When MODE is activated the UP arrow key scrolls through the various features and settings of the TI 007DL When the data log ging feature has been activated by pressing the MEM key the UP arrow is used to scroll through the various fields storage locations and fu...

Страница 4: ...8 8 INMM µs 1 8 8 8 8 INMM µs 1 8 8 8 8 INMM µs 8 0 APPENDIX B APPLICATION NOTES Measuring tubing When measuring a piece of tubing for wall thickness it may prove beneficial to have multiple delay lines with different radiuses for different tubing diameters The delay lines can be easily radiused by placing a piece of emery cloth around the tubing and moving the transducer back and forth until a ra...

Страница 5: ...lled couplant This fluid serves to couple or transfer the ultrasonic sound waves from the transducer into the material and back again Before attempting to make a measurement a small amount of couplant should be applied to the surface of the material being measured Typically a single droplet of couplant is sufficient After applying couplant press the transducer wearface down firmly against the area...

Страница 6: ... described in the following pages 15 6 0 TRANSDUCER SELECTION The TI 007DL is inherently capable of performing measurements on a wide range of materials from various metals to glass and plastics Different types of material how ever will require the use of different transducers Choosing the correct transducer for a job is critical to being able to easily perform accurate and reliable measurements T...

Страница 7: ...perator know the sound velocity of the mate rial to be measured A table of common materials and their sound velocities can be found in Appendix C 1 Make sure the TI 007DL is ON 2 Press the CAL key to enter calibration mode If the IN or MM symbol is flashing press the CAL key again so that the IN µs or M s symbols are flashing 3 Use the UP and DOWN arrow keys to adjust the displayed velocity up or ...

Страница 8: ...e front panel of the gauge and an audible indication beep will sound This improves the speed and efficiency of the inspection process by eliminating constant viewing of the actual read ing displayed The following procedure outlines how to enable and set up this feature Using the Beeper 1 Press ON OFF key to power up the unit 2 Press MODE key to activate features and settings 3 Press the UP and DOW...

Страница 9: ... a reading in the desired location The data logger will automatically advance to the next storage location in sequential order 9 Repeat step 8 as required Clearing a Storage Location The user may require a storage location that is currently full be over written This procedure is outlined in the following steps Note Assuming the steps in Using The Data Logger have been completed and step 8 is being...

Страница 10: ...splay will flash FILE F 01 or the last file used symbol Remember there are 10 files F 01 thru F 10 3 Press the SEND key to enter file setup The current file will be displayed F 01 F 03 etc 4 Press the UP DOWN arrow keys to scroll to the file 1 10 that will be used to record the measurements 11 4 0 MEASUREMENT MODES The TI 007DL is equipped with four measurement mode options echo echo interface ech...

Страница 11: ...TROMATIC ULTRASONIC MICROMETER MODEL TI 007DL OPERATING INSTRUCTIONS Distributed by ABQ Industrial LP USA Tel 1 281 516 9292 888 275 5772 eFax 1 866 234 0451 Web https www abqindustrial net E mail info abqindustrial net ...

Отзывы: