IMPORTANT:

When an ILR is found to be not working at all or not

working properly, see that the vaccine temperature is

within the recommended limits. 
Do not open the lid unless very essential.
Observe temperature from time to time and if you feel

that temperature may exceed the higher limit before

the ILR is repaired - transfer the vaccines to other

working ILR/refrigerator or cold-box.
If required use smaller pieces of commercial ice in

polythene bags in the cold-box - in absence of frozen

ice packs.
For any abnormal sound, smoke, smell etc. in the ILR

disconnect the plug from the wall socket and notify

refrigerator technician.

A)

THE ILR IS NOT WORKING

1.

Observe the `green' mains-on indicator light on

right on the front panel.

a) IF NOT GLOWING

I) Check if wall socket switch is OFF.
II) Check if the voltage stabilizer, if any, is working

or not.

III) Check if power supply is available at the wall

socket. This can be tested by connecting an

electric lamp or other appliances to the socket.

The lamp should glow or the appliances should

work, if supply is available at the socket.

If not available:
There may be some defect in the power supply circuit

(viz: blow-off fuse, loose connections, faulty switch or

socket, single phasing etc.) which should be rectified

through competent Electrical Maintenance

Technicians.
If power is available at the socket:

IV Check that the plug is inserted properly into the

socket.

V Check the plug for loose connection or

dislocation of power supply cable connections -

rectify if required.

VI)Remove the plug from socket. Check the

incoming and outgoing cable connections on the

terminal strip in the compressor compartment

(see page 6 & 7, actions 3B and 3C)

VII) Even after actions as above, the `yellow' lamp

is not glowing - Notify Refrigerator Technician.

b) IF GLOWING

I) Observe if the compressor cooling fan is

working. If the fan is working but the

compressor is not running - Notify Refrigerator

Technicians.

II) If the fan, too, is not working, check the

thermostat setting, if it is wrongly turned

completely anti-clockwise. If so, turn it

clockwise to a higher setting and observe if

the fan and the compressor are working now.

If working - Allow the ILR to run and adjust

thermostat to obtain recommended temperature.
If not working - Notify Refrigerator Technician.

B) COMPRESSOR IS WORKING, BUT

CABINET TEMPERATURE IS HIGHER

THAN NORMAL

NOTE: 

The cabinet temperature may go little higher than

normal, but should come down after some time

when:

New vaccines are stored
The lid is opened frequently or kept open for

longer duration
Unfrozen ice packs are put in for freezing. 

1.

Check if recommended space is left on back and

sides of the ILR for air circulation (see page 8,

action 6.

2. 

Check if there is too much frost formation in the

liner chamber on the inside walls. If so, defrost (see

page 13 for Defrosting and Cleaning)

3. 

Check if the compressor is cut `off' by the

thermostat before required temperature is attained.

If so, adjust thermostat setting.

7. TROUBLE SHOOTING

Содержание TCW 1990

Страница 1: ...ANUAL AL NO NOTICE D UTILIS TICE D UTILISA ATION TION MANU MANUAL DEL USU AL DEL USUARIO ARIO TCW 1990 ICE LINING REFRIGERATOR FREEZER GB REFRIGERATEUR CONGELATEUR A PAROI REFRIGERANTE ILR FRIGORIFICO CONGELADOR REVESTIDO DE HIELO F E ...

Страница 2: ...E LINING REFRIGERATOR FREEZER Handle Lid ice accu Rubber gasket Data label Thermometer 850 1050 630 1465 Compressor compartment Pilot lamp ON Thermostat for refrigerator function Drain plug Locking device Interior wall ...

Страница 3: ...ANING page 13 14 6 PERIODIC OBSERVATIONS CHECKS ACTIONS page 15 7 TROUBLE SHOOTING page 16 17 A ILR not working B Compressor is working but cabinet temperature is higher than normal C Cabinet temperature is lower than normal D Abnormal noise in the ILR 8 SHUTTING DOWN page 18 9 STORAGE OF VACCINES IN THE ILR page 18 INDEX 3 ...

Страница 4: ...age variations E g if the normal operating voltage is 230 volts 50 Hz A C it can work even when the supply voltage occasionally go as low as 150 volts or as high as 280 volts THERMOMETER To enable to measure the temperature inside the cabinet without opening the lid a thermometer is provided with its dial on the front side This makes monitoring of the vaccine temperature easy LOCKING SYSTEM The IL...

Страница 5: ...horities and keep the equipments repacked The ILR is to be installed in a well ventilated room avoiding direct sunshine or any other source of heat Power supply socket 15 amps socket with switch fuse and mains indicator lamp is advisable should be available nearest to the place of installation of the ILR 1 Unscrew and remove the screws holding the side cover of the compressor compartment on the ri...

Страница 6: ...ly for any damage dislocation or looseness etc of the components specially at locations A B C and D as below and rectify if necessary LOCATION A fig 4 Check that all 4 mounting brackets of the compressor are in place and secured 2 INSTALLATION 6 Fig 4 Fig 3 ...

Страница 7: ...ord to the terminal strip are tight LOCATION C fig 6 Check that the outgoing electrical connections from the terminal strip are tight LOCATION D fig 7 Check that the mounting brackets of the fan are tight and also that the fan blades are free to rotate 2 INSTALLATION 7 Fig 5 Fig 6 Fig 7 ...

Страница 8: ...ent space on all sides of the ILR keep it away from the walls for good circulation of air around it see below Back and left side should be minimum 3 cm away from the walls fig 9a Hinges can be pushed against the wall Right side should be minimum 10 cm away from the wall fig 9b 6 The ILR has 4 supporting legs If the ILR is not in a level and stable position adjust this by adding wooden blocks or si...

Страница 9: ...eads of the power supply cord to a suitable 3 pin plug as shown below fig 10 Lead of cord Pin of plug Green yellow Earth Blue Neutral Brown Phase CAUTION DO NOT CONNECT THE ILR TO POVER SUPPLY NOW 2 INSTALLATION Fig 10 ...

Страница 10: ...the 5 ice packs in the refrigeration compartment 5 Plug in the appliance for test The green lamp indicates that the unit is under power fig 13 6 Check in the compressor compartment that the compressor starts and the fan on the left side is running 7 Check that the blower in the bottom of the dividing wall is running Check by pushing with your fingers to the flap thermostat if a free moving to open...

Страница 11: ...ll all the 36 ice packs with water tight the cap carefully and stack them in the greater freezing compartment It is important that they are stacked exactly as shown on the sketch in order to obtain air circulation through the ice packs and in order to avoid damages caused by expansion during freezing fig 16 11 After correct installation of all the ice packs insert the plastic innerliner and fix it...

Страница 12: ...perature on the font thermometer is colder then the desired range from 2 to 8 C you must check if the water in the ballast container is frozen If this is the case you have to check the flap thermostat 3 When the temperature on the front thermometer is observed to be steady for several hours within the range from 2 to 8 C only then the vaccines to be preserved should be transferred to the ILR fig 1...

Страница 13: ... is not available the vaccines will have to be preserved in Cold Box or Vaccine carrier properly lining the same with frozen ice packs such that the vaccine temperature remains within the recommended storage temperature during the defrosting of the ILR DEFROSTING CLEANING PROCEDURE 1 When the frost formation around the evaporator is 6 to 10 mm thick it is time for Defrosting 2 Prepare the temporar...

Страница 14: ...apper the remove the ice formation around the surfaces never use a hairdryer to melt the ice formation 8 Before you insert the ice packs again you must be sure that all the inner parts are absolutely dry 9 Check that the blower and the flap thermostat are working correctly 10 If everything is 100 dry you can insert the plastic innerliner Don t forget to refit the drainage plug ...

Страница 15: ...ation has started around the evaporator and the ice bank Prepare for defrosting 2 Check if the gasket if not broken and if it seals tight against the frame 3 Check if the ballast container in the lid is not frozen MONTHLY 1 Clean the compressor compartment a Put off the switch if any and take out the plug from the wall socket b Take out the side cover of the compressor compartment see page 5 actio...

Страница 16: ... into the socket V Check the plug for loose connection or dislocation of power supply cable connections rectify if required VI Remove the plug from socket Check the incoming and outgoing cable connections on the terminal strip in the compressor compartment see page 6 7 actions 3B and 3C VII Even after actions as above the yellow lamp is not glowing Notify Refrigerator Technician b IF GLOWING I Obs...

Страница 17: ... floors c Keep some not frozen ice packs on and around the vaccines also d Run the ILR under surveillance and put it off when lowest recommended temperature is attained Put the ILR on again when the highest recommended temperature is reached and keep it running until the lowest recommended temperature is attained again Repeat as above till the ILR is repaired D ABNORMAL NOISE IN THE ILR In case of...

Страница 18: ...o a coil and bind and place it carefully to avoid damage 8 SHUTTING DOWN FOR STORING VACCINES 1 Keep the packets containing the vaccines in neat rows 2 Different vaccines should be kept separately to facilitate easy identification 3 Keep about 2 cm space between rows for circulation of air 4 Store DPT TT vaccines not touching the inside walls or bottom of the ILR It is better to keep them in the b...

Страница 19: ...19 ...

Отзывы: