By „Alarm-Set“ we mean the maximum and minimum admissible internal temperature. It can be modified as follows, according to
needs :
For /35-models :
Alarm COLD = minimum value set: -55°C , maximum value set: 5° below the „Temp-Set“.
Alarm WARM = maximum value set: -15°C, minimum value set: 5° above the „Temp-Set“.
For /86-models :
Alarm COLD = minimum value set: -95°C, maximum value set: 5° below the „Temp-Set“.
Alarm WARM = maximum value set: -65°C, minimum value set: 5° above the „Temp-Set“.
●
By pressing the relative WARM and COLD keys the „Alarm-Set“ values set can be detected. The relative „WARM-alarm“ or „COLD-
alarm“ LEDs light up while the set temperature is displayed. The following values have been preset:
For /35-models :
Alarm COLD : -40°C
Alarm WARM : -30°C
For /86-models :
Alarm COLD : -87°C
Alarm WARM : -77°C
●
In the setting of the „Alarm-Set“ proceed as follows:
●
Alarm-Cold = Press the “SET” and “COLD” key at the same time; the
“Alarm-Cold” LED lights up and the temperature changes in 1ºC stages
from the set value up to the end value. Once this value has been reached,
the display moves back to the original value. In order to set a value, sim-
ply release the keys once the temperature required is displayed.
●
Alarm-Warm = Press the “SET” and “WARM” key at the same time; the
“Alarm-Warm” LED lights up and the temperature changes in 1ºC stages
from the set value up to the end value. Once this value has been reached,
the display moves back to the original value. In order to set a value, sim-
ply release the keys once the temperature required is displayed.
●
By „Temp-Set“ we mean the desired internal temperature of the refrige-
rator. This temperature has been preset in the factory. If needed it can be
adjusted. Such adjustment must be done by a technician from the
Service Centre. If the „Temp-Set“ is changed, the „Alarm-Sets“ change
automatically.
●
In order to be able to detect the „Temp-Set“ set, press the „SET“ key. The
„Temp-Set“ LED will light up and the „Temp-Set“ value set will be dis-
played as long as the „SET“ key remains pressed.
●
Pre-set values:
For /35 appliances:
-35°C
For /86 appliances:
-82°C
T
EMPERATURE
D
ISPLAY
11
T
EMP
-S
ET
12
A
LARM
-S
ET
13
●
On the display screen the internal refrigerator temperature is displayed.
TEMP
ALARM
MEMORY
°C
SET
MEM
SET
-35.0
TEMP
ALARM
MEMORY
°C
SET
MEM
SET
-50.0
TEMP
ALARM
MEMORY
°C
SET
MEM
SET
-20.0
●
For /86 appliances:
If the temperature is colder than -99°C, the display
will appear;
temperatures warmer than -60°C will be shown with
.
II II
-- --
●
For /35 appliances:
If the temperature is colder than -60°C, the display
will appear;
temperatures warmer than -10°C will be shown with
.
II II
-- --
/86
/35
-60°C
-99°C
-10°C
-60°C
-- --
II II