SoilTest PRO Auto Soils Consolidator (ASC)
25-0500
Original Instructions
19
9901X0275 Issue 2
Default Sensor Calibration
WARNING
These default calibration values are only applicable to the default transducers supplied by
ELE International. The use of other transducers that do not have these characteristics may
result in damage to the machine. ELE International will accept no responsibility for damage
caused by using default parameters with different transducers.
Prior to calibration data from the DS8 application it is useful to use default calibration values
for the 15 kN and 15 mm sensors. This will be used to give an indication of the system
status.
Select the Calibration icon, enter the passcode (the default passcode is 937707) and
‘
Calibrate
’.
In turn select ASC Sensor 1 and 2. For ASC Sensor 1 and 2 select the DEFAULT option and
confirm. The parameters can then be viewed.
Sensor
Gain
Slope
Range (always S.I. Units)
1. Load Sensor
8
-13.75
15000 (N)
2. Displacement Sensor
9
0.0046
15 (mm)
NOTICE
The calibration data described below is not intended for the purposes of testing samples. It
is used as an indication of loading and is for safety purposes to prevent an overload
condition on the machine.