Whereas
•
GT = conductivity measured at temperature T (mS/cm)
•
GT
Ref
= conductivity (mS/cm) displayed by the instrument, calculated
back to the reference temperature T
Ref
•
α = linear temperature correction coefficient (%/°C); α = 0: no
temperature correction
•
T = measured temperature (°C)
•
T
Ref
= Reference temperature (20 °C or 25 °C)
-
After confirming the setup of the reference temperature, the α-coefficient
value is blinking.
1)
Set the α-coefficient value using ˄ and ˅.
2) Press
Read
to confirm.
3) Continue with TDS factor setting or press
Exit
to return to measurement
screen.
5.1.6 TDS factor
TDS is calculated by multiplying the conductivity value by the TDS factor.
-
After confirming the α-coefficient, the TDS value is blinking.
1) Set the TDS factor
using ˄ and ˅.
2) Press
Read
to confirm.
3) Continue with the Temperature unit setting or press
Exit
to return to
measurement screen.
5.1.7 Temperature unit
-
After confirming the TDS setting, the temperature unit is blinking.
1) Select the temperature unit ( °C or °F)
using ˄ and ˅.
2) Press
Read
to confirm and get back to the measurement screen.