Eldex Optos 1HM Скачать руководство пользователя страница 8

 

A Guided Tour of the Optos Metering Pumps 

Front Panel 

As you look at the front panel of your Optos Metering Pump (Figure 1.1), you will see: 

Component Details 

Local Interface: 

1.

 

Liquid Crystal Display (LCD).  

Displays flow rate, pressure (with optional pulse damper), other 

allowed functions.

 

2.

 

Up/Down Keys.  

UP

 and 

DOWN

 arrow keys

 

set flow rates, change pump settings.

 

3.

 

Permanent Functions Keys.  

DISPLAY

, Δ

MENU

, and 

RUN

/

STOP

 keys are control keys permanently 

defined for direct control of major instrument functions.

 

4.

 

Status LEDs.

  Indicate pump running or fault.

 

Pulse Damper:

  

optional pulse damper for minimized fluid delivery pulsation. 

5.

 

Outlet Port:

  Connection to downstream device. 

6.

 

Inlet Port:  

Connection from pump outlet. 

Liquid End Assembly: 

7.

 

Outlet Valve: 

 Provides tubing connection for pump output.   

8.

 

Cylinder:

  Contains pumping chamber and piston wash ports. 

9.

 

Inlet Valve: 

 Provides tubing connection to fluid source.  Inlet valve has an additional groove on 

the hexagonal portion of the valve. 

Rear Panel 

As you look at the rear panel of your Optos Metering Pump (Figure 1.2), you will see: 

Component Details 

1.

 

RS232:

  RJ11 connector for RS232 communication. 

2.

 

I/O:

  Input/output connections for remote analog control, pressure out, error out, remote start, stop. 

3.

 

Power Entry: 

 Provides power connection with modular cord. 

4.

 

Fuse:

  Housing for pump fuses. 

5.

 

On/Off:

  Turns the power to the pump on and off. 

Содержание Optos 1HM

Страница 1: ...ex Optos Metering Pumps Operator s Manual Eldex Laboratories Inc 30 Executive Court Napa CA 94558 Tel 707 224 8800 Fax 707 224 0688 www eldex com Rev Z 09112018 2018 Eldex Laboratories Inc All Rights...

Страница 2: ......

Страница 3: ...nections 10 Power Entry Module 10 Inputs Outputs 10 Plumbing 11 Solvent Preparation 11 Attaching Inlet Tubing 12 Priming The Pump 12 Attaching Outlet Tubing 12 Pulse Damper Use 12 Using the Piston Was...

Страница 4: ...19 Pump Limits 20 Viscosity Limits 20 Chemical Limits 20 Mechanical Limits 20 Temperature Limits 20 Shutdown Procedures 20 IV MAINTENANCE 21 Cleaning 21 Piston Seals 21 Replacing The Piston Seal 21 Re...

Страница 5: ...Return Procedure 34 Trademarks 34...

Страница 6: ...riate for the materials you use ATTENTION Gardez l esprit que nous avons test cette pompe avec le propanol 2 Les quantit s r siduelles du propanol 2 peuvent se trouver sur le dispositif de raccordemen...

Страница 7: ...t 25 psi is required on the outlet side of the pump The pump s flow rate is set directly in mL min using the up and down buttons on the user interface Flow rates are determined by piston displacement...

Страница 8: ...minimized fluid delivery pulsation 5 Outlet Port Connection to downstream device 6 Inlet Port Connection from pump outlet Liquid End Assembly 7 Outlet Valve Provides tubing connection for pump output...

Страница 9: ...3 Figure 1 1 Front View of Optos Figure 1 2 Rear View of Optos 1 2 3 4 5 6 7 8 9 1 2 3 4 5...

Страница 10: ...nges the display to different pump functions DISPLAY key toggles display back to main display of flow or flow and pressure with optional pulse damper RUN STOP key runs the pump and stops the pump Key...

Страница 11: ...own may vary from your specific configuration e g pressure is only displayed and high and low limit settings are only available on the Plus versions Basic Screens XX XXXmL XXXXpsi MAIN HI LIMIT XXXX H...

Страница 12: ...io View EPROM level Set remote control configuration Set piston materials of pump Send piston to forward position Set pressure limit ON OFF View stroke length View set voltage out Set home position of...

Страница 13: ...9 12 2HM 0 02 40 0 1500 250 250 2 mL 100 230V 50 60Hz 80 9 5 x 4 x 9 12 Stainless Steel Optos Model 3 Pump Specifications Model Flow Rate Range min to max mL min Max Pressure psi Piston Diameter inch...

Страница 14: ...x 4 x 9 12 2HI 0 02 40 0 1500 250 250 2 mL 100 230V 50 60Hz 80 9 5 x 4 x 9 12 PEEK Optos Model 3 Pump Specifications Model Flow Rate Range min to max mL min Max Pressure psi Piston Diameter inches Max...

Страница 15: ...9 5 x 4 x 9 12 2HMH 0 02 40 0 1500 250 250 2 mL 100 230V 50 60Hz 80 9 5 x 4 x 9 12 3HMH 0 04 80 0 750 250 500 4 mL 100 230V 50 60Hz 80 9 5 x 4 x 9 12 Maximum flow rate specifications are based on the...

Страница 16: ...ng a variety of fluids which require special handling procedures and safety precautions Consult the appropriate MSDS Material Safety Data Sheet supplied with the materials you will be using Be certain...

Страница 17: ...closure stops pump Outputs The Optos can send contact closures and an analog signal to remote devices The remote outputs available are defined in Table 2 2 Table 2 2 I O Position Label Function Descri...

Страница 18: ...air bubbles persist turn off the pump check the liquid level in the reservoir and the tightness of the inlet fitting If the inlet tubing is above the surface of the liquid or if the inlet fitting is...

Страница 19: ...of water with 20 methanol is suggested The pump cylinder has two barbed fittings installed Usually the lower fitting is used as the wash inlet and the upper fitting is used as the wash outlet Attach 1...

Страница 20: ...ll be activated and the stall stop condition The UP and DOWN keys change the setting of the various parameters When the setting is numeric the longer the key is held down the faster the value changes...

Страница 21: ...NU key from the main display Press the RUN key The motor will drive the piston to the home position To return to the main screen press the DISPLAY key Changing the Refill Rate Setup Optos pumps can be...

Страница 22: ...vision Level Optos pumps have EPROM software You can view the version of software installed To view the software version hold the DISPLAY key and press the MENU key from the main display or when in th...

Страница 23: ...ey from the main display or when in the advanced screens press the MENU key to display the limit screen Use the UP or DOWN arrow key to change the limit setting DIS limits de activated ENA limits acti...

Страница 24: ...he main display or when in the advanced screens press the MENU key to display the save state screen Use the UP or DOWN arrow key to change the limit setting OFF pump does not start pumping after power...

Страница 25: ...iew the stroke length of the piston of your pump by holding the MENU key and pressing the DISPLAY key from the main display or when in the advanced screens press the MENU key to display the stroke scr...

Страница 26: ...ex pumps are designed to operate under ambient temperature conditions If you take appropriate precautions to prevent condensation you can use the pump in a cold room Do not use the pump in an environm...

Страница 27: ...a leak may not always be visible particularly at low flow rates We suggest you place a wick or litmus paper up the drain slots just underneath the liquid end You should remove the wick or litmus paper...

Страница 28: ...n seal Any scratches or irregularities will require replacement of the liquid end 8 Clean all parts with a liquid detergent followed by a rinse Clean any fluid in the mechanical end of the pump 9 Inse...

Страница 29: ...rt the new valve into the liquid end assembly and tighten by hand until just finger tight Then use a wrench and tighten turn more You should tighten valves to the point at which no further leakage occ...

Страница 30: ...CTFE seal into place Slide the valve cartridge into the valve housing making certain it is oriented correctly Press another new CTFE seal into place The CTFE seal will extend approximately 0 020 0 03...

Страница 31: ...of mechanical assembly is piston broken pushrod frozen in bushing Are valves OK No Rebuild or replace valves excessive force used to tighten fittings to valves or valves to cylinder can cause sapphire...

Страница 32: ...fitting 1373 Inlet valve assembly cartridge style 1 8 Swagelok style fitting 1374 Outlet valve assembly cartridge style 1 8 Swagelok style fitting 1371 Inlet valve assembly cartridge style 10 32 fitt...

Страница 33: ...t valve assembly cartridge style 1 8 Swagelok style fitting 1374 Outlet valve assembly cartridge style 1 8 Swagelok style fitting 1371 Inlet valve assembly cartridge style 10 32 fitting 1370 Outlet va...

Страница 34: ...ludes two PTFE seals no tools 5946 Piston seal replacement kit includes five PTFE seals no tools Valves 1373 Inlet valve assembly cartridge style 1 8 Swagelok style fitting 1374 Outlet valve assembly...

Страница 35: ...ent kit includes five uhmw polyethylene seals Valves 1377 Inlet valve assembly cartridge style 1378 Outlet valve assembly cartridge style 1383 Valve cartridge kit rebuilds one inlet or one outlet valv...

Страница 36: ...includes five uhmw polyethylene metal free seals Valves 1377 Inlet valve assembly cartridge style 1378 Outlet valve assembly cartridge style 1383 Valve cartridge kit rebuilds one inlet or one outlet v...

Страница 37: ...lacement kit includes five uhmw polyethylene metal free seals and backup washers Valves 1377 Inlet valve assembly cartridge style 1378 Outlet valve assembly cartridge style 1383 Valve cartridge kit re...

Страница 38: ...retainer Piston not included Other Parts 6052 Retainer wash assembly For other parts consult factory For 1 4 Hastelloy C 276 Pumps Order No Description Piston Seals 6080 Piston seal assembly kit incl...

Страница 39: ...rate ID OKxyzaaa Reads ID of pump where x piston diameter 0 093 1 125 2 250 y piston stroke 0 125 1 250 2 500 z pump material 0 ss 1 pk Aaa EPROM revision RP OK xxxx 0 Pump specific Reads the pump pre...

Страница 40: ...ranties express or implied which extend beyond this description Eldex neither assumes nor authorizes any person to assume for it any other liability in connection with the sale and use of the products...

Страница 41: ......