5.3 The electronic unit is to be conveniently located in order to exclude the cable
tension when you plug to mains. The working mains voltage: ~ 220 V (±10%) or ~ 230 V
(- 10%, +6%), frequency 50 Hz.
5.4 When wiping the outer surfaces of the Device make sure that moisture does not
penetrate inside the electronic unit and coils-inductors. Protect the Device from dampness,
shock and impact.
5.5
CAUTION!
Do not use the Device with the damaged body, coils-indu
c
tors and
cables.
5.6 Be sure that the mains voltage is 220 V (±10%) or 230 V (-10%, +6%)
5.7 Safety precautions at therapeutic procedures:
a) the first treatment procedure time should not last longer than 20 minutes; at
treatment of two zones the total procedure time should not exceed 30 minutes,
b) the time of the first 3 procedure if the treated area is cervico-thoracic should be not
more than 10 minutes.
c) the direct application of the device on the heart region and brain is forbidden.
Note: there may be the signs of material outflow on the on the working surface of the
device body, which are not considered as the body damage and do not make influence on
the device serviceability .
CAUTION!
Do not lift or carry the device by the power cable.
Do not place the operative device close (less than 0,5 m) to magnetic data carriers,
audio, video-devices and others magneto-sensitive devices.
6 PREPARATION FOR USE
- After storage or transported at temperatures below +10 °C it should be kept in the
room at temperature (from +10 °C to +35 °C) for at least 4 hours;
- Before the application the outer surfaces of electronic unit and coils- inductors
should be disinfected by their wiping using a piece of fabric cloth watered with 3%
solution of hydrogen peroxide added with 0,5 % detergent of “Lotas” type or by 1%
solution of chloramine.
While cleaning avoid leakage of the disinfection solution or detergent inside electronic
unit and coils-inductors case.
To be sure in the presence of magnetic field, join the indicator of magnetic field to the
effective surface of each coil-inductors of operative device.
The common effective surface is the effective surface of all non twisted four coils.
At the beginning of the treatment procedure the patient takes comfortable position
which he could keep up to the end of procedure.